QbD1200+laboratory TOC analyzer
QbD1200+TOC detector can reduce interference between different sample measurements
The measurement results of QbD1200+TOC analyzer are consistent and have good repeatability
QbD1200+online TOC detector can support continuous monitoring of multiple different sensors
The QbD1200+TOC analyzer has reports for both self check upon startup and on-demand diagnosis
QbD1200+TOC analyzer also has over range recovery function
The QbD1200+TOC detector adopts a 10.4-inch color touch screen, with a more intuitive user interface
The QbD1200+TOC analyzer features a guided software interface for easy operation
The QbD1200+TOC detector comes with an automatic adjustment of measurement range/dilution function, which allows for measurement even without prior knowledge of sample concentration
QbD1200+laboratory TOC analyzer only requires one reagent to complete the corresponding measurement and analysis (self prepared or purchased from Hash)
The QbD1200+online TOC detector supports simple and convenient paperless functions, and can export data to network shared directories managed by Windows security policies
The reagent cost of QbD1200+TOC detector is low
No need for frequent maintenance, only once a year is required
No need to equip a separate computer
Calibration only takes 2 hours
The TOC detector is equipped with automatic system adaptability testing for USP and EP, as well as SDBS verification program for JP
The QbD1200+laboratory TOC analyzer complies with major pharmacopoeia regulations worldwide, including USP, JP, EP, IP, KP, and ICH
The simplified IQ/OQ/PQ certification procedure for QbD1200+TOC detector complies with ICH guidelines
The QbD1200+TOC analyzer adopts a simple automation process and can be applied in the following fields:
-USP/EP SST Bulk Water
-USP SST sterile water
-JP SDBS verification
-KHP calibration in accordance with JP standards
The QbD1200+TOC analyzer is mainly aimed at pharmaceutical companies' QC laboratories, drug research and development, drug testing institutes, power plants, CDC laboratories, third-party testing institutions, universities, and research institutes
The QbD1200+TOC detector can be widely used in industries such as pharmaceuticals, electricity, disease control, and tap water, meeting the strict requirements of the industry for water quality monitoring.
| technical indicators | |
|---|---|
| Performance Specifications | |
| measurement range | 0 ~ 100 ppm |
| precision | <2% or 3ppb (whichever is greater) |
| accuracy | ± 3% |
| calibration cycle | approximately 2 hours |
| analysis time | The fastest time is about 4 and a half minutes |
| Inorganic carbon treatment | |
| Do high content inorganic carbon samples require additional inorganic carbon elimination modules? | No |
| Is dynamic endpoint detection used to ensure complete removal of inorganic carbon? | is |
| The system's self diagnostic function ensures the integrity of data | |
| Will the instrument report TOC values if the following situations occur? | |
| The intensity of the ultraviolet lamp is too low | No |
| Insufficient carrier gas/empty storage tank | No |
| Insufficient reagents/bottle empty | No |
| Insufficient sample volume/small bottle empty | No |
| reagent | |
| Reagent supply | Purchase or self configure |
| Is there an option to prepare your own reagents | is |
| Total number of reagents (acid, oxidant, dilution water) | one |
| Measurement methods and guarantees | |
| oxidation | UV lamp+persulfate |
| Confirmation of UV lamp | Broadband silicate photodiode sensor |
| detector | Digital NDIR |
| Detector stability | Continuously verify according to the reference value |
| Detector moisture pre cleaning | Peltier cooler with continuous current monitoring |
| Distributor module | Using an injection pump, accurate to+/-1% of the volume |
| flow sensor | Dynamic fluid sensor (ensuring accuracy of sample and reagent volumes) |
| Selection of carrier gas | High purity air, oxygen or nitrogen |
| Confirmation of carrier gas pressure | Dual port pressure sensor |
| Closing of carrier gas valve | After the measurement is completed, the valve will close to avoid wasting gas |
| Automatic range function | have |
| Analysis mode | NPOC (Non Blowable Organic Carbon) |
| Data storage and retrieval | |
| Statutory authentication | 21 CFR, All measurement results in Part 11 are recorded in an encrypted database |
| Paperless report | Output report of network shared directory managed by Windows security policy through Ethernet |
| IO (Input/Output) | 3 USB ports, 1 Ethernet port |
| Data export format | PDF,CSV |
| user interface | |
| monitor | 10.4-inch high-resolution color touch screen |
| External PC | No external hardware required, embedded Windows 10 IoT Enterprise operating system |
| External keyboard/mouse | Optional, not mandatory |
| easy to operate | Intuitive user interface that guides operators through measurement settings for operation |
| Technical specifications | |
| calibration | Automated process: Perform 18 point calibration using KHP (5 concentrations, each repeated 3 times, 3 background) |
| Calibration solution | 5ppm C KHP (single 125ml bottle), diluted with reagents to concentrations of 1, 2, 3, 4, and 5ppm |
| Verification testing/system adaptability | Automated process: |
| USP SST Bulk Water Blank, 500ppb calculated as C: Sucrose, p-benzoquinone | |
| USP SST sterile water Blank, 8ppm as C: Sucrose, p-benzoquinone | |
| JP-16<2.59>verification Blank, 500 ppb calculated as C SDBS | |
| Compliance with regulations | |
| USP<643>(including sterile water SST), JP-16<2.59>, EP<2.2.44>, IP, CP, KP, US EPA 5310c | is |
| 21 CFR, 第11部分 | is |
| The ICH guidelines for instrument validation include: Accuracy, precision, detection limit, quantification limit, linearity, range, specificity, durability |
have |
| autosampler | |
| capacity | 64 sampling bottles (40mL glass vials TOC) |
| style | XYZ, Diaphragm perforation |
| Sample disk material | Aluminum anodizing |
| General technical data | |
| Instrument size | Width 320 x Depth 507 x Height 410 mm |
| power supply | 100-240 VAC,50/60 Hz,2A |
| weight | 14 kilograms (31 pounds) |
| Automatic sampler size | Width 366 x Depth 537 x Height 457mm |
| power supply | 100-240 VAC,47-63Hz |
| weight | 21 kilograms (45 pounds) |
| Working temperature and humidity | 5-35℃; Relative humidity up to 90%, non condensing |
| Unit used in the report | ug/L (ppb) |
| Order Information | |
|---|---|
| Item Number | description |
| analyzer | |
| 9880000 | QbD1200+laboratory TOC analyzer |
| 9890400 | AS-0640 automatic sampler (64 sample grades) |
| QbD1200+laboratory TOC analyzer&automatic sampler | |
| Accessory Information | |
|---|---|
| Item Number | description |
| Consumables and spare parts | |
| 9459400 | Concentrated reagents, QbD1200+ |
| 9459500 | 5 ppm C KHP calibration solution, 125ml |
| 9459600 | SDBS Verification Kit, 500 ppb C |
| 9459700 | USP System Applicability Kit (500 ppb) |
| 9459800 | USP System Applicability Kit (8 ppm) |
| 9459900 | Characteristic testing kit (1 ppm) |
| 9460000 | Stability testing kit |
| 9460100 | Verification Protocol Suite |
| 9891700 | Syringe replacement kit |
| 9890500 | Ozone processor replacement kit |
| 9898900 | Pipeline replacement kit |
| 9880200 | UV reactor replacement kit |
| 9890600 | Reagent bottle kit with customized bottle cap |
| 9898300 | QBD1200+power adapter |
| 9467200 | AS 0640 automatic sampler tray |
| 9454400 | AS 0640 Automatic Sample Injector No Flange Nut Expansion Tool |
| 9467400 | AS 0640 Automatic Sample Injector Needle Sleeve |
| SP6790 | AS 0640 automatic sampler diaphragm puncture probe |
| 2672700 | Sample bottle with diaphragm, 40 mL, TOC, 72 pieces |
| Hach Service Products | |
| SP-QLF1-QBD | QBD1200+Equipment Confirmation Service (IQ/OQ) |
| SP-COMF1-QBD | QBD1200+installation and debugging services |