GANIMEDE Total Phosphorus Analyzer
The GANIMEDE system adopts a highly integrated modular design, consisting of three core units working together:
After the user inputs the sample information, it commands the entire analysis process in a unified manner.
Equipped with a 36 bit (upgradable to 53 bit) sample tray, it enables automatic and continuous injection of bulk samples.
Automatically complete sample digestion and analysis.
The analysis process of GANIMEDEP total phosphorus analyzer is simple: the sample information is input into the control unit, and the automatic sampler automatically injects the sample into the digestion and analysis unit. The automatic needle flushing and needle sample mixing function ensure the accuracy of measurement results and pollution-free sampling.
1) The GANIMEDEP total phosphorus rapid analyzer has an automatic sampler with 36 sample bottles, and can also choose an automatic sampler with 53 sample bottles;
2) The GAIMEDEP total phosphorus analyzer can automatically clean the pipeline during each measurement gap to avoid the carrying effect of the pipeline;
3) Having wireless infrared communication connection and RS232 interface facilitates data download;
4) The GANIMEDEP total phosphorus analyzer uses pre made reagents, which can greatly shorten the preparation time of the reagents;
5) The pipeline of the GANIMEDE total phosphorus analyzer analysis system allows particles with a particle size of 100 μ m to pass through;
6) The testing of total phosphorus does not require sample preparation, the use of fume hoods, or exposure to toxic and harmful reagents, making the entire analysis process time-saving and safe;
7) The same controller can manage both GAIMEDEP and GAIMEDEN analysis systems simultaneously.
The GANIMEDE total phosphorus analyzer is suitable for water quality analysis in industrial and municipal laboratories, large sewage treatment plants, power industry, and other fields.
| technical indicators | |
|---|---|
| Analysis method: | 880nm ammonium molybdate colorimetry |
| Dissolving method: | Acid persulfate rapid digestion method (150 ℃, 6bar) |
| Sample pH: | pH 2-10 |
| Measurement range: | 0.01-3.8mg/L PO4-P |
| Accuracy: | <2% at 1mg/L |
| accuracy: | <1% |
| Analysis unit calibration method: | Single point or multi-point calibration |
| Dissolve and analyze unit pipelines: | Allow particles with a diameter of 100 to pass through |
| Automatic sampler turntable: | With a diameter of 240mm, there are two models: (1) with 36 sample bottles and stirring function; (2) with 53 sample bottles and no stirring function |
| Control unit display screen: | color screen display |
| Weight of digestion and analysis unit: | 40 kilograms |