The precious metal element analyzer is specifically designed for precious metal testing, using technology to analyze X-ray fluorescence spectra for non-destructive and rapid analysis of precious metal composition. The instrument uses Si PIN detector and basic parameter method, which can accurately and without error analyze other small amounts of impurity elements contained in materials such as gold and platinum. Its stability and reliability are at the highest level in the industry.
1、 Precious metal element analyzerProduct Information
This instrument is specifically designed for precious metal testing, using technology to analyze X-ray fluorescence spectra for non-destructive and rapid analysis of precious metal components. The instrument uses a Si PIN detector with a resolution of 149 ± 5eV and is used in the basic parameter method to accurately analyze other small amounts of impurity elements contained in materials such as gold and platinum. Its stability and reliability are at the industry level.

2、 Precious metal element analyzerApplication fields:
Testing of the production, sales, and recycling of products related to precious metals such as Pt, Au, Pd, Ag, Rh, etc;
Used for composition analysis of various metal alloys (copper alloys, iron alloys, stainless steel, tin alloys, etc.).
IIIFunction:
Open work curve allows for the establishment of new workspaces based on demand
Combination of empirical coefficient method and basic parameter method
Test samples: Various shapes of samples to meet the testing needs of large samples
Safety performance: Maze like design, * achieving radiation safety indicators
Simple testing operation, implementing a foolproof operation mode
Diversified report templates to meet various testing needs
4Instrument parameters:
Analysis software: Ux-6280 precious metal analysis software
Analysis of element types: 26 elements can be analyzed simultaneously
Analysis method: Basic parameter method (quantitative)
Analysis element range: Na-U, capable of simultaneously analyzing 26 elements
Measurement range: ppm-100%
Measurement time: 100-300s
Accuracy: 0.1% (constant relative error)
Resolution: 149 ± 5eV
Pipe pressure: 5KV-50KV
Pipe flow: 50uA -1000uA
Power: 350W
Dimensions: 640 (W) * 474 (D) * 412 (H) (mm3)
Sample compartment size: 300 (W) * 300 (D) * 100 (H) (mm3)
Weight: Approximately 48Kg
Working environment: 15-30 ° C, relative humidity ≤ 85% (non condensing)
Working power supply: AC220-240V, 50/60Hz, 350W
5Standard configuration:
Detector: 2-level semiconductor electric refrigeration; The thickness of the detector crystal is 500 μ m; Detection sensitive area 6mm2; Be window thickness of 1mil
High voltage power supply: High voltage range 0-50kV, power 50W, 8h stability 0.05%
X-ray tube: thin beryllium window side window X-ray tube (Mo target, Rh target)
High definition camera device
Accurate positioning device
Automatic lid opening device