E-mail
eric@sdicelc.com
-
Phone
18911571363
-
Address
15th Floor, Building 6, Courtyard 2, Wuliqiao Second Street, Chaoyang District, Beijing
Beijing Zhongqing Guotou Environmental Protection Technology Co., Ltd
eric@sdicelc.com
18911571363
15th Floor, Building 6, Courtyard 2, Wuliqiao Second Street, Chaoyang District, Beijing
-The working principle of Polymetron 9611sc online phosphorus meter and phosphate online analyzer is as follows:
Low range phosphate meter: The low range phosphate meter is an online continuous monitoring instrument that uses ascorbic acid colorimetric method to determine phosphate concentration at a wavelength of 880nm.
High range phosphate meter: The high range phosphate meter is an online continuous monitoring instrument that uses the vanadium molybdate colorimetric method to determine the concentration of orthophosphate at a wavelength of 480nm.
-Application industry:
● Online continuous monitoring of phosphate content
● Steam drum boiler water in thermal power plants
● Other industrial water systems
-Instrument features:
Low maintenance volume and low downtime rate
Continuous operation for 90 days, achieving unmanned operation
● Time saving maintenance
● Avoiding downtime due to malfunctions
● Clean, fast, and easy reagent replacement
● Grab in and Grab out functions
Equipped with advanced Prognosys pre alarm system, it can diagnose measurement and maintenance parameters to help you make proactive arrangements for maintenance operations.
| Polymetron 9611sc online phosphate meter, phosphate online analyzer technical specifications | |
|---|---|
| Measurement range: | Low range: 4-5000 μ g/L, calculated as PO4; |
| High range: 200-50 μ g/L, calculated as PO4 | |
| accuracy: | Low range: ± 4 μ g/L or ± 4% of the reading, whichever is greater; |
| High range: ± 500 μ g/L or ± 5% of the reading, whichever is greater; | |
| Repeatability: | Low range: ± 1% |
| High range: ± 500 μ g/L or ± 5% reading, whichever is greater | |
| Detection limit: | Low range: 4 μ g/L |
| High range: 200 μ g/L | |
| Response time: | Generally, there is a slight change with temperature variation for 9 minutes (at a temperature of 25 ° C) |
| Reagent consumption: | Each reagent consumes 2 liters every 90 days, with a cycle time of 15 minutes |
| Sample pressure: | 2 - 87 psi |
| Sample temperature: | 5 - 50 °C (41 - 122 °F) |
| Water sample flow rate: | 55 - 300 mL/min |
| Channel Number: | 1, 2, 4, programmable |
| Sampling: | Manual sampling function (Grab in) |
| Manual sampling function (Grab out) | |
| Installation method: | Wall mounted, panel mounted, or desktop mounted installation |
| Pipe diameter size: | Sampling tube and sample bypass drainage tube 6mm (¼ - in.) |
| Air purification inlet: 6mm (1/4 in.) | |
| Chemical drain pipe: 9.5 mm (3/8-in.) | |
| Compatible controller: | SC200,SC1000 |
| Power requirements: | 100 - 240 V AC 50/60 Hz |
| Power output: | 4 - 20 mA |
| Dimensions (height x width x depth): | 804 mm x 452 mm x 360 mm |
| Weight: | 20 kg without reagents, 36.3kg with reagents |
| Protection level: | IP56 |
| ordering guide | |
|---|---|
| 9611.KTO.PH.A1U | Hash 9611sc high range phosphorus analyzer, AC, 1 channel, containing hash reagents for 90 days of use. |
| 9611.KTO.PH.A2U | Hash 9611sc high range phosphorus analyzer, AC, 2 channels, including hash reagents for 90 days of use. |
| 9611.KTO.PH.A4U | Hash 9611sc high range phosphorus analyzer, AC, 4 channels, including hash reagents for 90 days of use. |
| 9611.KTO.PL.A1U | Hash 9611sc low range phosphorus analyzer, AC, 1 channel, containing hash reagents for 90 days of use. |
| 9611.KTO.PL.A2U | Hash 9611sc low range phosphorus analyzer, AC, 2 channels, including hash reagents for 90 days of use. |
| 9611.KTO.PL.A4U | Hash 9611sc low range phosphorus analyzer, AC, 4 channels, including hash reagents for 90 days of use. |
| Polymetron 9611sc online phosphate meter, phosphate online analyzer optional accessories | |
|---|---|
| 2036100-CN | Low concentration phosphate determination reagent kit, including: reagents 1-3, standard solution 1-2 |
| 2036700-CN | High concentration phosphate determination reagent kit, including: reagents 1-3, standard solution 1 |
| 6788307 | Maintenance kit, for low concentration phosphate determination, 1 channel |
| 6788308 | Maintenance kit, for low concentration phosphate determination, 2/4 channels |
| 6788309 | Maintenance kit, for high concentration phosphate determination, 1 channel |
| 6788310 | Maintenance kit, for high concentration phosphate determination, 2/4 channels |