E-mail
eric@sdicelc.com
-
Phone
18911571363
-
Address
15th Floor, Building 6, Courtyard 2, Wuliqiao Second Street, Chaoyang District, Beijing
Beijing Zhongqing Guotou Environmental Protection Technology Co., Ltd
eric@sdicelc.com
18911571363
15th Floor, Building 6, Courtyard 2, Wuliqiao Second Street, Chaoyang District, Beijing
characteristic
Four models
- CarboVis®:CODsensor
- NitraVis®Nitrous nitrogen sensor
- NiCaVis®:COD\\Nitrous nitrogen sensor
- NiCaVis®NI:COD\\Nitrate nitrogen\\Nitrous nitrogen sensor
on-line measurementCOD、BOD、TOC、TSSAnd nitrate nitrogen, nitrite nitrogen
The patented optical system ensures accuracy, stability, and reliability
Full spectrum assurance results and compensation
Ultrasonic cleaning system ensures maintenance
Titanium alloy andPEEKMaterial, more resistant to chemical corrosion
Standard plastic cushioning protective cap, impact resistant, super durable
overview
The new spectral sensor has a diameter of60mmThe sensor contains a set of precise spectral analysis components, which can accurately analyze the organic pollution concentration at the inlet, biochemical tank or discharge outlet of municipal sewage plants in the ultraviolet visible bandCOD、BOD、TOC、DOC、SACIt can also be testedTSSTotal suspended solids concentration, nitrate nitrogen, and nitrite nitrogen concentration. Easy to use, no need for reagents, no need for sample pretreatment.
The patented optical system design ensures measurement accuracy, stability, and reliability, while the drift free sensor ensures reliable data even after prolonged use without the need for routine maintenance. Using a complete full spectrum length range from200nmarrive720nmperhaps200-390nmCompared to the single wavelength or multi wavelength principle, it has better measurement results and compensation for interfering materials such as turbidity. The built-in ultrasonic cleaning system reduces manual window cleaning time and can also add compressed air cleaning for special harsh environmental applications. Using high-tech materials-Titanium alloy andPEEkIt can be used in very harsh environments and has a long service life.
measuring principle
The basic working principle is UV visible continuous spectral scanning. When a beam of light passes through the test liquid, any substance in the liquid will absorb the light, reducing the intensity of the transmitted light. The instrument measures the absorbance at different wavelengths to depict the absorption curve. By analyzing the absorption curve, the test component can be qualitatively and quantitatively tested. At the same time, solid particles in the liquid will scatter light in various directions, causing a decrease in the intensity of transmitted light (at all wavelengths). Continuous spectral scanning can effectively compensate for the influence of changes in the background composition of the tested liquid. The probe tested using this method has a certain testing interval and cannot be tested continuously.
WTWThe spectral sensor adopts an optical design, with a symmetrical measurement system and reference system, the same optical elements, light intensity, light quantity, and optical path length, so it has automatic drift compensation and does not require conventional zeroing.
NitraVis /CarboVis /NiCaVisOptical sensors can automatically adjust the flicker frequency of the light source according to different samples, ensuring that the light signal is neither too strong nor too weak, with a very low signal-to-noise ratio and reliable measurement data.
Reliable measurement
UV/VISsensorCODtotalAnd the laboratoryCODComparison Chart
UV/VISsensorCODtotalAnd the laboratoryCODComparison Chart
Automatic cleaning
NitraVis /CarboVis /NiCaVisOptical sensors have multiple cleaning methods, including ultrasonic cleaning or compressed air cleaning, as well as composite mode ultrasonic cleaning+Compressed air cleaning. The sensor will automatically select the cleaning cycle. If you choose to blow with air, you need to set the corresponding relay function.
Ultrasonic cleaning system
The probe is equipped with an ultrasonic cleaning system as standard, which ensures long-term reliable operation and saves maintenance. Optical lenses will vibrate slightly under the action of ultrasound, which can effectively prevent pollutants from adhering to the optical lenses, greatly improving the stability and reliability of testing.
Air blowing device (optional)
When the on-site conditions are extremely harsh, it is recommended that users choose another set of high-pressure air blowing device to achieve dual safety and ensure the reliable operation of the system.
structure
Refer to the figure below, from the light source1The emitted light passes through a lens2After diffusion, it enters the measuring cell in the form of parallel light3, then through a converging lens4 After focusing, it reaches the detector5Generate sensing signals. In the figure,6To measure the signal, there is a test liquid passing through it,7As reference signals, without testing the liquid, these two sets of signals are received by photodiodes and converted into current signals, which are then converted into the required component concentrations through certain algebraic operations.
External dimensions
Multiparameter measurement
Spectral sensors can provide multi parameter measurement of carbon
at most5Carbon parameters can be measured, displayed, and transmitted simultaneously,
COD total/ COD soluble/ TOC/ BOD/ DOC/ SAC/ SAC soluble/ UVT254Choose five out of eight
increase254nmPenetration measurement(UVT254)
CODThe solubility algorithm can be used forWWTPApplication in Aeration Tank(Like denitrification/Methanol control
Technical Parameter
|
model |
NitraVis |
NitraVis |
CarboVis |
CarboVis |
NiCaVis |
NiCaVis |
NiCaVis |
|
|
|
measuring principle |
UV-VIScontinuous spectrum200-720nm |
UVcontinuous spectrum200-390nm |
|
||||||
|
application |
Discharge port |
Entrance, aeration tank |
Discharge port |
Entrance, aeration tank, and discharge outlet |
Discharge port |
Entrance, aeration tank, and discharge outlet |
Discharge port |
|
|
|
measuring range [mg/l] |
NO3-N: |
NO3-N: |
COD:
BOD:
TOC:
SAC: |
COD:
BOD:
TOC:
SAC: |
COD:
BOD:
TOC:
SAC:
NO3-N: |
NO2-N: 0.00-75.00mg/L NO3-N: 0.00-150.0mg/l COD:0-20000mg/L TOC:0-20000mg/L SAC:0-5000m DOC:0-12500mg/L BOD:0-8000mg/L UVT:0.00-100%
|
NO2-N: 0.00-25.00mg/L NO3-N: 0.00 -50mg/L COD: 0.0-800.0mg/L TOC:0.0-500.0mg/L SAC:0.00-600.0m DCO:0.0-500.0mg/L BOD:0.0-500.0mg/L UVT:0.00-100% |
|
|
|
TSS |
Entrance, aeration tank: |
Discharge port: |
entrance:
Discharge port: |
entrance:
Discharge port: |
|
|
|
|
|
|
Mechanical Construction |
Shell: Titanium alloy3.7035、PEEK;Lens: Sapphire |
||||||||
|
pHrange |
pH 4… pH 12 |
||||||||
|
velocity of flow |
≤3rice/second |
||||||||
|
Maximum pressure |
less than1 hope earnestly |
||||||||
|
Immersion depth |
At least the measuring tank should be submerged in the water sample, at most10 Depth in meters |
||||||||
|
Position orientation |
arbitrary |
||||||||
|
size |
802 × 59.9mm(Length)XDiameter) |
||||||||
|
weight |
make an appointment4Kg |
||||||||
|
power consumption |
8W |
||||||||
|
Warranty period |
Two years |
||||||||
|
Ordering Information |
||
|
model |
describe |
order No |
|
NitraVis ® 701 IQ |
Nitrous nitrogen UV visible spectroscopy probe with built-in ultrasonic cleaning |
481 044 |
|
NitraVis ® 701 IQ TSdouble probe |
Same as above, suspended solid concentration can also be tested |
481 045 |
|
NitraVis ® 705 IQ |
Nitrous nitrogen UV visible spectroscopy probe with built-in ultrasonic cleaning |
481 046 |
|
NitraVis ® 705 IQ TSdouble probe |
Same as above, suspended solid concentration can also be tested |
481 047 |
|
CarboVis ® 701 IQ |
COD/BOD/TOC/DOC/SACSpectral probe with built-in ultrasonic cleaning |
481 048 |
|
CarboVis ® 701 IQ TSdouble probe |
Same as above, suspended solid concentration can also be tested |
481 049 |
|
CarboVis ® 705 IQ |
COD/BOD/TOC/DOC/SACSpectral probe with built-in ultrasonic cleaning |
481 050 |
|
CarboVis ® 705 IQ TSdouble probe |
Same as above, suspended solid concentration can also be tested |
481 051 |
|
NiCaVis ® 705 IQ double probe |
COD/BOD/TOC/DOC/SAC/NO3Spectral probe with built-in ultrasonic cleaning |
481 052 |
|
NiCaVis ® 701 IQ NI |
UVSensors are used to measure nitrite, nitrateCOD、TOC、BOD、DOC、SACandUVT254 |
481054 |
|
NiCaVis ® 705 IQ NI |
UVSensors are used to measure nitrite, nitrateCOD、TOC、BOD、DOC、SACandUVT254 |
481055 |