Plastic tensile testing machine (small door machine) testing machine standard: HG/T 2369-1992 Technical conditions for rubber plastic tensile testing machine.
Equipment namePlastic tensile machine (small door type machine)
Model: CMT4104
Equipment structure and working principle:
This device adopts a single space structure, with a space type of lower space, mainly composed of a host (including rack, base, transmission system), fixtures, deformation measurement system, software system, and electrical system.
Equipment meets the following standards:
1. Testing machine standard: HG/T 2369-1992 Technical conditions for rubber and plastic tensile testing machine
2. Test method standard:
GB/T 1040.1-2018 Determination of Tensile Properties of Plastics Part 1: General Principles
GB/T 1040.2-2022 Determination of Tensile Properties of Plastics Part 2: Test Conditions for Molding and Extrusion Plastics
GB/T 1040.3 Determination of Tensile Properties of Plastics Part 3: Test Conditions for Thin Films and Sheets
Plastic tensile machine (small door type machine)Technical Specifications:
1. Maximum test force: 10kN;
2. Accuracy level of testing machine: 0.5 level;
3. Test force measurement range: 0.4% -100% FS, no grading throughout the entire process;
4. Relative error of test force indication: within ± 0.5% of the indication;
5. Experimental resolution: 1/1000000 of the maximum experimental force (with constant resolution throughout the entire process);
6. Deformation measurement range: 0.2% -100% FS;
7. Relative error of deformation indication: within ± 0.5% of the indication;
8. Deformation resolution: 1/± 500000 of the maximum deformation amount;
9. Relative error of displacement indication: within ± 0.5% of the indication;
10. Displacement resolution: 0.04 µ m, up to 0.01 µ m;
11. Range of force control rate adjustment: 0.005-5% FS/S;
12. Relative error of force control rate: within ± 0.5% of the set value;
13. Deformation rate adjustment range: 0.02-5% FS/S;
14. Relative error of deformation control rate: When the rate is less than 0.05% FS, it is within ± 1% of the set value; When the rate is ≥ 0.05% FS, it is within ± 0.5% of the set value;
15. Beam speed adjustment range: 0.001-500mm/min;
16. Relative error of crossbeam speed: within ± 1.0% of the set value when the speed is less than 0.01mm/min; When the speed is ≥ 0.01mm/min, within ± 0.5% of the set value;
17. Constant force, constant deformation, constant displacement control range: 0.5% -100% FS;
18. Constant force, constant deformation, and constant displacement control accuracy: When the set value is less than 10% FS, it is within ± 1% of the set value; When the set value is ≥ 10% FS, it is within ± 0.5% of the set value;
19. Beam movement stroke (excluding fixtures): 1100mm;
20. Effective test width: 400mm;
21. Overall dimensions of the host (length x width x height): 700 x 580 x 1920 (mm);
22. Power supply: 220V/50Hz ± 10%, 750W;
23. Host weight: 250kg.