Welcome Customer !

Membership

Help

Shenzhen Yixinfeng Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shenzhen Yixinfeng Technology Co., Ltd

  • E-mail

    exinfeng01@163.com

  • Phone

    13926596645

  • Address

    Jinhai Building, No. 109, Songgang Section, Guangshen Road, Songgang Street, Bao'an District, Shenzhen

Contact Now

Pesticide residue detection GC9310-VI gas chromatograph

NegotiableUpdate on 01/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
According to General Rule 2341 Pesticide Residue Determination Method, gas chromatography is used to determine the residual levels of some pesticides in medicinal herbs, decoction pieces, and formulations. The specific configuration is as follows: Pesticide residue detection GC9310-VI gas chromatograph
Product Details

Pesticide residue detection GC9310-VI gas chromatograph



According to General Rule 2341 for the determination of pesticide residues, usinggas chromatographDetermine the residual levels of some pesticides in medicinal herbs, decoction pieces, and preparations. The specific configuration is as follows:

gas chromatograph1 set: including 1 host, 1 set of electronic capture detector (ECD), 1 set of hydrogen flame detector (FID), 2 sets of capillary injectors, and 1 set of packed column injection. Including installation kit and accessories

1 set of chromatography workstation software: with audit tracking function

One capillary chromatography column: pesticide residue column, ON-1,30m*0.32mm*0.5um

One hydrogen generator: SPH-300

One air generator: SPA-2000

1 bottle of high-purity nitrogen: purity 99.999%, including pressure reducing valve

Computer printer: 1 set

Shenzhen Yixin Instrument Equipment Co., Ltd. provides thegas chromatographOn site installation, debugging, and training.

Pesticide residue detection GC9310-VI gas chromatograph
Parameters and performance:

Main performance

GC9310-VIgas chromatograph(220V,50Hz)

6-way heating system

10 file libraries

16th order programmed heating column box

Vacuum fluorescent screen display function, easy to operate. Chinese and English can be switched freely

Multiple program functions (ignition, memory, self diagnosis, time program, automatic rear door opening, etc.), reverse control function (various parameters can be remotely operated), and ten work file operation functions.

Multiple detector selection and configuration functions (FID, TCD, FPD, ECD).

Ethernet interface

Host and detector parameters

Host size: 660 * 480 * 480mm Net weight: 55kg

Power supply: 220V ± 10% (50Hz ± 0.5Hz)

Power: ≤ 2000W Environmental temperature: -5 ℃~35 ℃

Relative humidity: ≤ 85%

ECD detection limit: Mt ≤ 2 × 10-13 g/s (sample: n-hexane σ, -666);

Linear range: ≥ 10 3

Baseline noise: ≤ 10 μ A

FID detection limit: ≤ 3 × 10-12g/s

Baseline noise: ≤ 5 × 10-14A

Baseline drift: ≤ 2 × 10-13A/30min


GC9310-VIgas chromatographDetermine the standard spectra for detecting 9 types of organochlorine pesticide residues:

Analysis conditions:

Injector temperature: 280 ℃ Carrier gas: nitrogen

Detector temperature: 300 ℃

柱箱温度:100℃ 柱流量: 1.5ml/min

Program heating conditions: 10 ℃/min -220 ℃ for 12 minutes, then increase to 250 ℃ at a rate of 8 ℃/min