Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

PTC-660 temperature calibrator (dry furnace) overview description details technical specifications description PTC-660 temperature range: 33 to 660 ° C (91 to 1220 ° F) To the top of the page details 1. High precision

Product Details

PTC-660 temperature calibrator (dry furnace)

describe

PTC-660 temperature range: 33 to 660 ° C (91 to 1220 ° F)


To the top of the page

details

1.High precision, with a maximum accuracy of± 0.06 ° C (± 0.11 ° F), using 4-wire resistance true value measurement technology

2.Excellent stability:0.01˚C

3.Better temperature consistency, the unique dual zone heating well ensures excellent temperature field consistency

4.Intelligent reference probe,The JOFRA reference probe is equipped with an intelligent connector and a built-in chip that stores calibration data (temperature correction parameters) and other information. This is a true plug and play calibration system

5.USB communication, all PTC type dry body furnaces can connect and communicate with computers through USB interfaces

6.Efficiency improvement, fast heating capability/Refrigeration effectively improves work efficiency. Rising to 350 ℃ only takes 7 minutes

7.Calibration work order function, using the work order function can achieve fully automatic calibration



To the top of the page

TECHNICAL INDEX

1.Temperature range:33660℃;

2.Accuracy:±0.3(Built in standards,Type A, 33~420 ℃±0.5(Built in standards,Type A, 420~660 ℃),±0.15℃(join in marriageExternal standardOnlyBC type);

3.Stability:±0.04℃;

4.Radial consistency:0.1℃;

5.Display resolution:1/Adjustable at 0.1/0.01 ℃;

6.Calibration chamber (calibration sleeve) specifications(mm):Φ25× L160

7.Heating time(room temperatureTo the highest temperature:20minute;

8.Cooling time (from maximum temperature to)room temperature):51minute;

9. Automatic stepping: maximum20 steps, adjustable holding time for each point, calibration results can be stored;

10.Maximum power:1150W;

11.Instrument size (length)X Width x Height):362×171×363mm;

12. Weight:8.9kg。

The following parameters are limited totype B

13.Current measuring range/Accuracy: 0~24mA/±(0.02%Rdg+0.01%F.S);

14.RTD resistance measurement range/accuracy: 0~4000Ω/±(0.005%Rdg+0.005%F.S);

15.RTDtemperaturemeasuring range/Accuracy: Pt1001000/±0.04@0℃(Pt100)±0.06@350℃(Pt100)

16.thermocoupleMV measurement range/accuracy: ± 78mV/± (0.0)2%Rdg+0.01%F.S);

17.Thermocouple temperature measurement type/Accuracy: E/J/K/T/R/S/B/N ± 022@155℃(K)±0.32@660℃(K)

18.Cold end compensation range/Accuracy: 0-40 ℃/± 0.35

19.Switch test: dry contact5V,2.5mA;

20.24V circuit power supply: 24VDC ± 10%, maximum 28mA.


To the top of the page