- Phone
-
Address
Shanghai Pujiang Town High tech Park Caohejing Development Zone
Shanghai Jingpai Instrument Co., Ltd
Shanghai Pujiang Town High tech Park Caohejing Development Zone
PRP horizontal rotatorStandard Operating Procedure (SOP)
Scope of application: This laboratory uses a specialized digital horizontal rotary oscillator for syphilis non-specific antibody detection (TRUST/RPR)
1. Purpose
Standardize the operation, use, maintenance, and upkeep of the digital syphilis rotary oscillator to ensure uniform and stable test conditions, and to guarantee accurate, reliable, and traceable results of syphilis serological agglutination tests.
2. Scope of application
Applicable to the entire process of using a digital syphilis rotary oscillator for TRUST, RPR and other non-specific syphilis antibody detection by all testing personnel in this laboratory.
3. Responsibilities
3.1 Operator: Strictly follow this SOP to complete instrument operation, recording, and daily cleaning.
3.2 Department Head/Quality Manager: Supervise the implementation of SOP, review instrument usage and maintenance records.
3.3 Equipment administrator: responsible for regular calibration of instruments, fault reporting, and file management.
4. Instrument information
• Instrument name: Digital syphilis rotary oscillator
• Oscillation mode: horizontal circular rotation
Standard speed: 100 ± 2 r/min
Standard time: 8 minutes
• Storage environment: Room temperature 18-28 ℃, relative humidity ≤ 70%, no strong light, no vibration, no corrosive gas
5. Preparation before operation
5.1 Environmental Inspection
The instrument is placed on a stable and horizontal experimental table, with no debris, no standing water, and no interference sources around it.
5.2 Instrument Inspection
Check that the power cord is securely connected, the casing is not damaged, the control panel displays normally, the tray is not loose, and the protective cover is intact. Start up and run for 30 seconds without load, confirm that there is no abnormal noise or severe shaking.
5.3 Item Preparation
• Serum sample to be tested (centrifuged, no hemolysis, no lipid blood, no clots)
TRUST/RPR kit (antigen equilibration to room temperature)
Specialized reaction plate, micro sampler, suction head
• Timer, record sheet, biosafety protective equipment
5.4 Quality Control Requirements
Positive and negative controls must be set up synchronously for each batch of testing, and only after passing the quality control can the results be interpreted.
Model: Jipad IVG
Product Name:PRP horizontal rotator
Manufacturer: Shanghai Laboratory Instrument Factory
Product Overview:
1. Used for syphilis detection, screening and testing of syphilis RPR, TRUST, VDRL, etc., for qualitative and semi quantitative detection of non-specific antibodies against Treponema pallidum.
2. The horizontal sample tray should be equipped with a clamp groove to fix the reaction plate and avoid slipping out of the horizontal rotator during the experiment.
3. Complies with the health industry standard WS/T491-2024 "Guidelines for Non Specific Antibody Testing of Syphilis".
Technical parameters:
1. Operation mode: Horizontal circumference
2. Turnover diameter: 19mm
3. Allowable load capacity: 1kg
4. Speed range: 20-230rpm
5. Time setting range: 0-999 minutes
6. Operation mode: timed/continuous, the number of mixing time can be displayed in reverse, and a warning will sound when the mixing time is up. If not timed again, the machine will automatically stop. The buzzer and countdown numbers display dual prompts.
7. Input voltage (frequency): AC100-240V (50/60Hz)
8. Input power: 10W
9. Working face size: 310 × 219mm (ordinary type)
10. Allowable ambient temperature: 5-50 ℃
11. Allowable relative humidity: 80%
12. Dimensions: 320 × 315 × 120mm
13. Scheduled batch settings: 1-3 (cyclic)
14. There is a fixed transparent cover, and there are moisturizing devices on both sides of the cover with sponge dishes for constant temperature and humidity.
15. Three combination reaction programs can be set, and one click reaction program button design:
oneGroup: RPR, TRUST experimental reaction program: rotation speed of 100 ± 2r/min; Timed for 8 minutes
Group 2: Reaction program for VDRL-CSF specimens: rotation speed of 180 ± 2r/min; Timed for 8 minutes
The third group: VDRL - reaction program for serum samples: rotation speed of 180 ± 2r/min; Timed for 4 minutes
Model: Jipad IVXPRP horizontal rotator
Manufacturer: Shanghai Laboratory Instrument Factory
Product Overview:
1. The digital syphilis level rotation instrument is applied to promote whole blood plasma serum antigen antibody reaction, syphilis non-specific examination, syphilis detection, syphilis RPR, TRUST, VDRL screening and detection tests, etc. It is used for qualitative and semi quantitative detection of syphilis Treponema pallidum non-specific antibodies.
2. The horizontal rotation speed should comply with 100r/min ± 2 r/min (RPR) or 180r/min ± 2r/min (VDRL).
3. The time control accuracy meets the requirement of ± 1s/min.
4. The digital electronic control components can preset fixed reaction programs for RPR/TRUST and VDRL tests.
5. The horizontal sample tray should be equipped with a clamp groove to fix the reaction plate and avoid slipping out of the horizontal rotator during the experiment.
6. The speed and time adjustment methods should be button type, displayed in digital status,quietType, digital inspection time is more accurate.
Technical parameters:
1. Operation mode: Horizontal circumference
2. Turnover diameter: 19mm (optional 22mm)
3. Allowable load capacity: 1kg
4. Speed range: 20-230rpm
5. Time setting range: 0-999 minutes
6. Operation mode: timed/continuous
7. Input voltage (frequency): AC100-240V (50/60Hz)
8. Input power: 10W
9. Working face size: 310 × 219mm (ordinary type)
10. Allowable ambient temperature: 5-50 ℃
11. Allowable relative humidity: 80%
12. Dimensions: 320 × 315 × 120mm
13. Net weight: 4.6Kg
15. Digital syphilis serum culture shaker with transparent cover (optional)
Model: Jipad IIA
NamePRP horizontal rotator
Product Description:
The Shanghai Laboratory Instrument Factory brand digital mixer is used for screening syphilis PRP, TRUST, VDRL, USR, TPPA detection, PRP testing paper and testing experiments, as well as qualitative and semi quantitative detection of non-specific antibodies against Treponema pallidum. It can also be applied to mixing various types of petri dishes, enzyme-linked immunosorbent assay plates and other flat test vessels. It has two modes of continuous constant speed operation and timed constant speed operation.
Main features:
The upgraded product of the traditional RPR detection card oscillator can also be used for mixing various flat experimental vessels such as culture dishes and enzyme-linked immunosorbent assay plates, with a wide range of applications.
Compact and stable structure, quiet and smooth operation, no noise.
Using high-quality DC deceleration motors with an effective lifespan of over 1000 hours.
Electronic speed and time adjustment for easy operation.
The LED displays the speed and time in digital format, with a countdown function that automatically stops after the countdown ends. It has two modes of continuous constant speed operation and timed constant speed operation.
Technical parameters:
1. Operation mode: Circular
2. Turnover diameter 19mm ± 1
3. Allowable load-bearing capacity: ≤ 1kg
4. Speed range: 20-230rpm
5. Speed control accuracy: ± 3rpm
6. Speed/timer display: LED
7. Time setting range: 1-999min
8. Operation mode: timed/continuous
9. Allowable ambient temperature: 5-40 ℃
10. Allowable relative humidity: 80%
11. IP protection level: IP21
12. Power supply: AC100-240V 50/60Hz
13. Instrument input power: 10W
14. Working face size: 305 × 215mm
15. Volume: 320 × 325 × 110mm
16. Net weight: 4.2Kg
Syphilis rotary oscillatorPRP horizontal rotatorDigital rotary mixer, mixer, horizontal oscillator, buzzer alarm and digital countdown digital display dual prompts, with card slot, buzzer and countdown digital display dual prompts.