Advantages: One click fixed reaction program with three groups
Product Name:Syphilis digital mixer
Model: Jipad IVG
Manufacturer: Shanghai Jingpai Instrument Co., Ltd
Jipad IVG mixer and Jipad IVX mixer, differences: 1. Timing setting batch: 1-3 (cyclic), RPR/TRUST experimental reaction program, VDRL-CSF specimen reaction program, VDRL serum specimen reaction program. Difference 2: Equipped with a moisturizing box, placed with a moisturizing sponge, or added with cotton strips.
Syphilis digital mixerProduct Overview:
1. Used for syphilis detection, screening and testing of syphilis RPR, TRUST, VDRL, etc., for the non-specific detection of Treponema pallidum
Qualitative and semi quantitative detection of heterologous antibodies.
2. The horizontal rotation speed meets the requirements of 100r/min ± 2 r/min (RPR) or 180r/min ± 2r/min (VDRL)
3. The time control accuracy meets the requirements, and the time can reach 15 minutes.
4. The digital electronic control components can preset fixed reaction programs for RPR/TRUST and VDRL tests.
5. The horizontal sample tray should be equipped with a clamp groove to fix the reaction plate and avoid slipping out of the horizontal rotator during the experiment.
6. Comply with the health industry standard WS/T491-2024 "Guidelines for Non Specific Antibody Testing of Syphilis" and the one click reaction program button
Design.
Syphilis digital mixerTechnical parameters:
1. Operation mode: Horizontal circumference
2. Turnover diameter: 19mm
3. Allowable load capacity: 1kg
4. Speed range: 20-230rpm
5. Time setting range: 0-999 minutes
6. Operation mode: timed/continuous, the number of mixing time can be displayed in reverse, and a beep will sound when the mixing time is up. If not timed again,
Automatic shutdown. The buzzer and countdown numbers display dual prompts.
7. Input voltage (frequency): AC100-240V (50/60Hz)
8. Input power: 10W
9. Working face size: 310 × 219mm (ordinary type)
10. Allowable ambient temperature: 5-50 ℃
11. Allowable relative humidity: 80%
12. Dimensions: 320 × 315 × 120mm
13. Scheduled batch settings: 1-3 (cyclic)
14. There is a fixed transparent cover, and there are moisturizing devices on both sides of the cover with sponge dishes for constant temperature and humidity.
15. Three combination reaction program can be set
Di group: RPR, TRUST experimental reaction program: rotation speed of 100 ± 2r/min; Timed for 8 minutes
Group 2: Reaction program for VDRL-CSF specimens: rotation speed of 180 ± 2r/min; Timed for 8 minutes
The third group: VDRL - reaction program for serum samples: rotation speed of 180 ± 2r/min; Timed for 4 minutes
Digital mixer: It has a mixing timing setting and can be operated in a loop. After placing the test cards in batches, press the timing button each time to perform a timed control mixing operation. The internal microcomputer has a memory function for the mixing setting time. Cards with the same mixer time can be mixed without repeated settings each time. The number of mixing time can be displayed in reverse, and a warning will sound when the mixing time is up. The last timer will automatically stop the machine. Equipped with an electronic daily clock, the digital display can convert date or time at any time Detachable work disk. Used for detecting syphilis RPR, TRUST, etc.
Jipad IVG timed batch settings: 1-3 (cyclic), RPR/TRUST experimental reaction program, VDRL-CSF specimen reaction program, VDRL serum specimen reaction program. At least Shanghai Jingpai Instrument's digital syphilis rotary oscillator has 3 batches of mixing timing settings, which can be operated cyclically. After placing test cards in batches, simply press the timing button each time to perform timed mixing control.
SOP Manual for the Use and Operation of Digital Horizontal Rotating Instrument
Syphilis digital mixerUsage:
1. Open the packaging, remove the instrument and accessories, plug in the power cord, and connect the instrument to the power source.
2. ① Timed and constant speed operation: Press the corresponding button at the "time" position, "▲" for increasing time, "▼" for decreasing time, long press for continuous setting, select to the desired running time (time setting value displayed on the left LED); Press the corresponding button at the "speed" position again, "▲" for increasing speed, "▼" for decreasing speed, long press for continuous setting, select to the desired operating speed (the speed setting value is displayed on the right LED); Press the "○" button ("Start/Stop" button) to start the instrument running.
② Continuous constant speed operation: Select the timing as "000" and then choose the desired operation time
Press the "○" key to achieve continuous constant speed operation.
○ 3 Fixed working mode: Press the button for the desired test experiment name, then press the "○" button, and the instrument will start running.
If a pause is required during the operation of the instrument, press the "○" key to stop the instrument.
4. Set batches at regular intervals: 1-3 (cyclic), RPR/TRUST experimental reaction program, VDRL-CSF specimen reaction program, VDRL serum specimen reaction program. There are at least 3 batches of mixing timing settings, which can be operated in a loop. After placing the test cards in batches, simply press the timing button each time to perform timed mixing control.
● Precautions
When selecting an instrument working platform, please try to choose one with a high friction coefficient and a horizontal coefficient
Higher platform.
2. This instrument is a precision electronic product, and should avoid severe vibration during operation and transportation.
3. Keep the original packaging of the instrument for return to the factory for repair and use.
4. Before sending the instrument for repair, please clean it thoroughly to avoid contamination by hazardous substances. Please place the instrument away
Send back to the original packaging box.
When the instrument is not in use for a long time, please disconnect the power supply and place the instrument in a dry, clean place
In a stable environment at room temperature.
6. The AC power supply used for the instrument should have a reliable grounding wire, and the power supply voltage should be within the specified range
Within the scope.
When changing the position of the work board to clean the instrument plane, the switch should be turned on and manual operation is prohibited,
Prevent damage to parts.
● Product List
1. One user manual 2. One instrument 3. One power cord 4. Transparent cover (optional)
● Purpose and characteristics:
First of all, thank you for choosing the Jipad IVG digital level rotation instrument. This instrument is used for non-specific antibody detection of syphilis and is designed specifically for VDRL, RPR, TRUST experiments, in compliance with the WS/T491 health industry standard. It has the advantages of reasonable structure, novel design, complete functions, convenient operation, digital display and control speed and time. It has two modes of continuous constant speed operation and timed constant speed operation.