PHCBI 530L vertical high-pressure steam sterilizer

NegotiableUpdate on 12/12
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The PHCBI 530L vertical high-pressure steam sterilizer is equipped with a leakage protection function, which automatically cuts off the power supply in case of leakage.
Product Details

PHCBI 530L vertical high-pressure steam sterilizerparameter

PHCBI 530L vertical high-pressure steam sterilizerMLS-530L specification parameters

External dimensions (W * D * H)

520*660*881(mm)

Inner diameter of the cavity (diameter * depth)

φ370*470mm, Convenient access to items, clean the height inside the chamber

External material

ABS resin high-strength exterior cover, double-sided galvanized steel plate, surface spray treatment with strong corrosion resistance

Tank material

3mm thick corrosion-resistant stainless steel, surface mirror treatment to increase heat radiation effect and improve heating rate

Tank lid mechanism

Manual up and down opening and closing, safety locking mechanism, automatic locking during equipment operation

Steam water connection plate

Prevent steam droplets from leaking into the box when opening the lid

effective volume

50L

weight

About 105kg

device switch

Equipped with leakage protection function, it automatically cuts off the power supply in case of leakage

exhaust valve

Two solenoid valves: one for full opening and one for slow opening, to reduce liquid splashing caused by pressure during liquid sterilization

safety valve

Spring fully open safety valve, automatically operates when the maximum pressure exceeds 0.255Mpa, in compliance with YY 0154 standard

sensor interface

Internal temperature sensor interface, pressure gauge interface, can output pressure and temperature simultaneously, ensuring the correspondence between pressure and temperature. Sample temperature sensor interface (optional)

cooling method

You can choose between natural cooling and forced cooling fan cooling, with forced cooling time reduced by more than half compared to other models. Even in other modes, as long as

When the saturated steam temperature is below -2 ℃, the fan blades will move and reach 60 ℃

The fan stops running. The forced cooling button can be set at any time before or during operation. 121 ℃ -100 ℃ only takes 3 minutes

pressure gauge

0~0.4Mpa, Accuracy level 1.6, installed in front for easy confirmation

Maximum operating pressure

0.255Mpa

Operation control panel

7-inch ultra large touch screen, even wearing gloves, can easily set temperature and time, and display operation functions

power supply

AC220V 50HZ,11.5A

environmental conditions

Temperature: 5-40 ℃, relative humidity: ≤ 85%, air pressure: 70kPa~

106kpa, Below an altitude of 3000m, it meets the standards of the medical device industry

Built in cooling water tank

Large capacity 5L water tank reduces the number of cooling water treatments. The auxiliary muffler reduces the noise during exhaust. The external observation window can confirm the water level at any time

Internal water injection volume

≥6.5L, Support long-term sterilization and multiple sterilization

Anti blocking function

The bottom drain of the container is equipped with a filter as standard to prevent foreign objects and impurities from entering and blocking

caster

4, with locking function

performance parameters

Heater and sensor protection measures

Place a heating baffle on the ground inside the cylinder to protect the heater and sensor. Maximum water level, water level indication

Heater type

Stainless steel heating tube 1000W × 2

Heater circuit control

Solid state relay, pulse control output

sensor

Pt100 thermistor

temperature range

Equipment, liquid sterilization engineering: 105 ℃~135 ℃

Dissolving Engineering: 60 ℃~110 ℃

Insulation engineering: 45 ℃~60 ℃

Preheating temperature: 45 ℃~80 ℃

temperature control

Microcomputer PID Control

Setting and display mode

Use the up and down keys to set the numerical display mode, with a minimum resolution of 0.1 ℃

Appointment Function

1. Time mode: Set in units of 1 minute within the range of 00:00 to 23:59, default to the current day.

2. Date and Time Mode: Set the unit to year, month, day, and hour, within the range of 00:00 to 23:59, with a resolution of 1 minute.

Timing and timing decomposition function

1. Set time: 1 minute to 99 hours and 59 minutes, resolution: 1 minute

2. Set time: 100 hours~999 hours, resolution: 1 hour

Key lock function

Lock and release keys on keyboards other than the 'Start/Stop' key in maintenance mode to prevent accidental operation during operation

preheating function

Preheat 45 ℃~80 ℃ in advance, reducing sterilization delay time by 10-20 minutes

memory function

Each mode has three memory units that can store and remember temperature, time, forced cooling, etc. Up to 12 memory programs can be saved, and each program can be renamed

Fault recording function

20 faults and their types can be recorded based on year, month, and day

Accumulated sterilization/frequency/time

Users can check and grasp the operation status of the device at any time

Temperature measurement function for sterilized items

Temperature measurement, sterilization, and dissolution time timing using item sensors (optional)

Data transmission function

Timing output, alarm output, RS485 communication function, temperature output

Alarm and safety parameters

alert content

Sensor abnormality, relay short circuit, heater disconnection, lid lock abnormality, drainage switch abnormality, sample sensor disconnection, exhaust valve abnormality, air burning abnormality, temperature rise abnormality, safety valve abnormality.

Accessories and Optional Parts

accessory

2 φ 332 × D195.5 baskets, 1m silicone drain pipe, 1 set of sterilization test cards (30 sheets), 1 steam water cup, 1 cooling water tank, 1 heater baffle, 1 filter plug, and 1 brake

Optional accessories

Item temperature sensor, can temperature testing sensor, recorder, temperature output terminal (4-20mA), external alarm output terminal, time reaching output terminal, standard basket, raised basket

other

industry standard

The cylinder body manufacturing complies with the GB150-2011 standard.

Product Execution Standard: YY/T 1007-2018

Manufacturer's warranty

1.5 years after equipment installation

Calibration service, 3Q certification

Having professional qualifications for calibration and 3Q certification

Application for Pressure Vessel Use License

Corresponding to the application standards for pressure vessels in various regions