1PET performance sensitivity and scattering fraction detection moduleDF-200A Product Introduction and Structure
Used for quality control testing of PET performance in PET-CT, PET-MR and other equipment, in compliance with the WS 817-2023 Positron Emission Tomography (PET) Equipment Quality Control Testing Standard.
The coordinates of the capillary hole are (0,1cm), (0,10cm), (10cm, 0); The coordinates of the sensitivity module socket are (0,0), (0,10cm); Adjustable horizontal design.
1) Mass<10kg;
2) Electric control (non pneumatic);
3) Peristaltic pump method for packaging;
4) Three channel automatic control of the nuclide to be packaged;
5) Automatic control of packaging process;
6) The control software can display real-time readings of the activity meter.
The nuclear medicine image quality control analysis software can analyze DI COM format images, calculate intrinsic spatial uniformity, intrinsic spatial linearity, spatial resolution, system sensitivity, and PET/CT image fusion accuracy.
Spatial resolution; Sensitivity; Scattering fraction; NECR; We can provide technical training services for quality control and testing of nuclear medicine equipment. Accuracy: 1; Measurement range: 1-11.
70cm in length; Wall thickness 1mm; Comes standard with 2 wire source tubes.
Schematic diagram of PET sensitivity module
70cm in length; 20cm in diameter; comes standard with 2 wire source tubes.
Schematic diagram of scattering fraction model
The nuclide used for spatial resolution detection is 18F, and its specific activity is recommended by the manufacturer.
The test source is a high specific activity radioactive point source placed inside a capillary glass tube, with a line diameter of less than 1 mm in any direction. The point source is arranged in the following six positions:
a) Parallel PET longitudinal axis at 1/2 of the center of the field of view and 1 cm at the center of the cross-section;
b) Parallel PET longitudinal axis at 1/2 of the center of the field of view and 10 cm at the center of the cross-section;
c) Parallel PET longitudinal axis at 1/2 of the center of the field of view and at 20 cm of the cross-sectional center;
d) Parallel PET axial field of view, 3/8 from the center and 1 cm from the cross-sectional center;
e) Parallel PET axial field of view, 3/8 from the center and 10 cm from the cross-sectional center;
F) Parallel PET long axis axial field of view, 3/8 from the center and 20 cm from the cross-sectional center.
Figure A.4 Schematic diagram of spatial resolution monitoring point source support
2PET performance sensitivity and scattering fraction detection moduleDF-200A detection performance
PET phantom multifunctional bracket: can support sensitivity phantom and capillary measurement of corresponding PET performance indicators; Adjustable horizontal design to meet the precise positioning of CT laser lights in X, Y, and Z directions.
Meet the quality control requirements of PET standard WS817-2023 and NEMA-2018 for testing deviations of 1 cm, 10 cm, and 20cm from the center.
Meet the quality control requirements for (1,0), (10,0), and (20,0) specified in PET standards WS817-2023 and NEMA-2012.
Horizontal adjustment in both forward and reverse directions, suitable for various machine detection conditions, ensuring that the test module is parallel to the center of the PET field of view.