1、 Product Introduction
PAV aging tester+vacuum degassing boxThe Pressure Aging Vessel (PAV) can simulate the oxidative aging of asphalt during road use and is suitable for evaluating the antioxidant aging ability of different asphalt under test temperature and pressure conditions.
2、 According to the standard
The requirements specified in T0630-2025 "Asphalt Pressure Heating Aging Test PAV Method" in JTG 3410-2025 "Test Code for Asphalt and Asphalt Mixtures in Highway Engineering", SH/T 0774-2005 "Asphalt Accelerated Aging Test Method (PAV Method)" in the petrochemical industry standard of the People's Republic of China, and reference to test methods such as ASTM D6521 and AASHTO-R-28
IIIPAV aging tester+vacuum degassing boxTechnical specifications
(1) PAV-1-1 Asphalt Pressure Aging Chamber
1. Power supply
Power supply voltage: AC220V ± 10% 50Hz
Rated power: 1100W
Rated current of fuse: 10A
2. Pressure vessel
Working volume: 10L
Working medium: Air
Rated working temperature: 110 ℃± 0.5 ℃
Rated working pressure: 2.1MPa
Design pressure: 2.2 Mpa
Hydraulic test pressure: 2.75 MPa
Safety valve opening pressure: 2.15 Mpa
Safe bursting disc bursting pressure: 2.6 MPa
3. Pressure controller
Accuracy: ± 0.01MPa
4. Temperature and time controller
Accuracy: ± 0.1 ℃
Time slot: 64x1080min
Power off hold function: Yes
5. External dimensions:840×520560Mm (length x width x height)
6. Control mode: Fully automatic touch screen control
(2) PAV-1-2 Asphalt Aging Vacuum Box
1. Power supply
Power supply voltage: AC220V ± 10% 50Hz
Rated power: 900W
Rated current of fuse: 10A
2. Vacuum container
Working volume: 6.5L
Working medium: Air
Rated working temperature: 170 ± 0.5 ℃
Rated working pressure: pressure 15kPa
Pressure control method: touch screen control
3. Temperature and time controller
Temperature control accuracy: ± 0.1 ℃
Time control method: touch screen
4. External dimensions:440×520×540Mm (length x width x height)
4、 Installation and Debugging
1. Open box inspection
Before installing the instrument, please carefully count the instrument and accessories according to the purchase agreement and packing list.
2. Installation environment requirements
The instrument should be installed in a ventilated, bright, and dry indoor environment. The height of the table on which the instrument is placed should be 70cm-90cm, and the table should be made of flame-retardant materials.
The instrument should be placed steadily, with all four feet evenly stressed and the end face kept horizontal.
3. Connect power supply
The power supply voltage is AC 220V, and the voltage is stable within ± 10%. The form and size of the power socket should match the plug configured for the instrument.
Insert the power plug into the power socket, the power indicator light will turn on, press the switch, and the control instrument will start working.
If the power is turned on and the indicator lights on and off the instrument panel do not light up, the first thing to check is whether the plug and socket are loose and whether the power is cut off.