LSP-2000Ophthalmic Lens Optical Analyzerbased onXia Ke-Hartmann(Shack-Hartmann)Optical principles can be used to measure the optical parameters of ophthalmic lenses in air or solution. The measured objects include contact lenses(RGPSoft contact lenses, corneal reshaping lenses, scleral lenses, etc., as well as various intraocular lenses with spherical, multifocal, multifocal, and wavefront guided designs (such asICLTheIOLEtc.).
LSP-2000Ophthalmic Lens Optical Analyzercomply withGB/T 11471.5andISO 18369-3The relevant content about the Hartmann method tester is one of the two recommended devices for measuring contact lenses using the solution method.
LSP-2000Ophthalmic Lens Optical AnalyzerIt is a high-level metrology solution for measuring contact lenses and intraocular lenses in air and liquid. Its modular design allows you to configure accurately according to your own needs and expand as your requirements grow in the future.
Technical Specifications:
technology |
l Based on Schalke-Hartmann's wavefront sensor and image processing technology l Measuring ophthalmic lenses in air or solution |
function |
l Measure spherical diopter, cylindrical diopter, cylindrical axis position, prism, prism angle, additive power, optical power topographic map, wavefront aberration(Zernike)Wait. l Automatically measure lens diameter and ellipticity. l Automatic detection of axis markings on cylindrical mirrorsnote1 |
software version |
SHSWorks PRO; SHSWorks autoCL; triple illumination(Triple lighting optional) |
wavelength |
540 nm ± 10 nm |
Sample testing platform |
Fixed platform/manualYplatform/manualX-Yplatform/electricYplatform |
Sphericity |
- 35D … +35Din the airnote2,7 |
Prism degree |
highest10Din the airnote3 |
field of view |
l Optical power measurement:8.5 mmnote4 l Lens image detection:20 mmnote4 |
horizontal resolution |
Optical power measurement:91 x 91Measure the dot matrix (default) or157 x 157Measurement lattice (optional) Lens image:2Million pixels (default) or9Million pixels (optional) |
Optical power reproducibility |
< 0.02 D(1 secThe actual test results show that the lens has been movednote5 |
Optical power repeatability |
< 0.002 D(1 secThe actual test results show that the lens has not movednote5 |
Uncertainty of optical power |
< 0.03 Dnote5 |
Measurement duration |
< 0.2~1Seconds (including data collection and evaluation)note6And the results show that |
size/weight |
≈ 260 × 524 × 560 mm³ (W×D×H) / 17 kg |
computer |
includeWindows 10 LTSC 64Position (English version) or updated version |
document |
CECertificates, calibration certificates, user manuals, etc |
including attachments |
l Check the operation with tweezers and tools l dust cover |
note1The program can automatically recognize common identifiers, and special identifiers can be added as needed.
note2When using the solution method for measurement, this optical power measurement range will cover-100D … +100DThe prescription prescription degree depends on the refractive index of the lens.
note3The difference in optical power between the highest and weakest meridians.
note4:± 3 %.
note5Use calibration lenses in the air for operation.
note6:autoCLActivated,ZernikeZernike reconstruction, refraction data evaluation
note7The calibrated aperture of this system is4.5mm


LSP-2000Version difference table:
project |
LSP-2000Basic version |
LSP-2000premium version |
Lens platform |
manualX-Yplatform |
electricYplatform |
Open measuring slot |
none |
have |
Measurement slot lighting |
none |
have |
Constant temperature filtration circulation pump |
none |
have |
The remaining functions and accessories are the same. If you need higher configuration or customization, please further communicate with us.