Online magnesium ion automatic titration analyzerKM-MG400
Magnesium ions are involved in water-Rock reactions, high-temperature evaporation, biological metabolism, and electrochemical processes are all key indicators in the industry, and their monitoring is directly related to safety, energy efficiency, and compliance costs. Suggest selecting matching detection technologies based on industry characteristics and establishing a real-time warning mechanism.
This deviceTo meet the needs of on-site online analysis, advanced optical titration detection methods are adopted, which have the advantages of miniaturization, integration, and automation.
Measurement method: complexometric titration(EDTA)
reference standard
Technical Specification for Flue Gas Desulfurization Engineering in Thermal Power Plants - Limestone/lime-Gypsum MethodHJ 562-2010
Limestone for Thermal Power Plants-Water Quality Control Standards for Wet Gypsum Desulfurization WastewaterDL/T 997-2006
Determination of Calcium and Magnesium in Water QualityEDTATitration methodGB/T 7477-1987
Industrial Circulating Cooling Water Quality StandardsGB/T 500502007
How to translateSanitary Standards for Drinking Water
GB5749-2006How to translateApplication Fields: Surface Water and Groundwater/Chlor alkali chemical salt water
/desulfurization wastewater / Monitoring of Industrial Water UseSelectable display units for switching:
mg/lmmol/lM
mol/l
Measurement principle: Magnesium ion online analyzer is an online continuous analysis instrument that uses
LED
Automatic complexometric titration analysis using a light source to measure magnesium ion concentration.
The typical analysis program of an analyzer is set inside the system, such as rinsing the optical reaction cell with a sample, extracting a portion of the sample, adding one or more reagents, such as buffer or masking agent, and then performing the first measurement as a reference measurement value. Reference measurement values can eliminate interference factors such as sample color and turbidity, as well as various colors generated from reagents and refracted light.
Using standard titrants, step by step precise addition to the magnetic stirring measuring cell, the titration is finally completed, and the sample color instantly changes from pink to blue. Artificial intelligence algorithms are used to determine the end of titration.
Features and advantages
The national standard method has accurate measurement, good repeatability, stability and reliability;
Equipped with automatic cleaning function, no need for regular calibration;
Comes with a circulation cup, automatically samples and measures from the circulation cup, without affecting the flow of soft water. There are no restrictions on water supply flow and pressure, making it easy to install and highly adaptable;
Large scale matching automatic dilution module;
Equipped with advanced modular configuration, it has a small size and is easy to install. Connect the medication, sampling interface, and wastewater pipeline to operate; Has an alarm indication for missing sample input; Long term automatic operation, low maintenance, and low operating costs;Measurement can be triggered by signals from the upper computer; 4A sturdy and sturdy casing designed specifically for industrial environments, ensuring optimal online operation and fully protecting electrical, mechanical, and liquid pipeline components;
Analysis time:5-15min; Use of medication:
Seed medicine
Sampling frequency (cycle time): 5-120min
, adjustable settings;technical indicatorsMeasurement method: complexometric titration(Color comparison)Range:10-500ppmThe20-1000ppm
、100-5000
The1000-20000ppm
Accuracy: 5%F.S.resolution0.1ppmpower supply220VAC,
12VDC The24VDC
optionalInstallation method: wall installation or workbench supportOperating Temperature5-60°C
Protection levelIP65 ,ABSSize:6500Lx600Hx230mm
weight: about kilogramAnalog output: One channel4 – 20mA
Digital output:1RS485
,MODBUS RTU
Alarm relay:Road relay (over limit alarm), can freely set alarm limitsSample temperature:
5°-65°C
Recommended pressure for imported samples:2kg
belowExport sample pressure: atmospheric pressure (no back pressure)Recommended sampling flow rate for rapid circulation system:
1Connection: Sampling to fast loop system with flexible entrance, entrance 16mm
2Export requires no back pressure. 4supporting list
3Online hardness tester 1platform
4Reagents 6Bottle(
5Measure the flow pool 1a
6Outer diameter of pipeline 1rice
7Sampling stainless steel valve 1a
8Wiring terminal block 1a
9Cleaning agents and syringes 1set
10User Manual 1share
11Certificate of Conformity 1share
Product nameplate
aManual measuring tools set
Other optional accessories:1 7Metal chassis10Stainless steel or carbon steel spray coating
2touch screen Inch orInch true color
3multi-channel
2-8通道
4Multi level filtration
5Reagent constant temperature
6