Welcome Customer !

Membership

Help

Gopu Instrument (Shanghai) Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Gopu Instrument (Shanghai) Co., Ltd

  • E-mail

  • Phone

  • Address

    5th Floor, Building 5, 599 Huiwang East Road, Jiading District, Shanghai

Contact Now

Online industrial water residual chlorine/chlorine dioxide analyzer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Online industrial water residual chlorine/chlorine dioxide analyzer: mainly composed of a controller matched with residual chlorine sensors Bsens650 (platinum electrode) or Bsens660 (gold electrode) and BAF615 residual chlorine dedicated constant flow channel. It can monitor residual chlorine, chlorine dioxide, and ozone in water, using the constant voltage principle for stable and reliable measurement, and easy installation. Mainly used for drinking water, hospital wastewater, secondary water supply, industrial process water, etc.

Product Details

Online industrial water residual chlorine/chlorine dioxide analyzer

Model: PM8202CL

在线工业用水余氯/二氧化氯分析仪

Online industrial water residual chlorine/chlorine dioxide analyzer

generalization:Mainly composed of a controller paired with residual chlorine sensors Bsens650 (platinum electrode) or Bsens660 (gold electrode) and BAF615 residual chlorine dedicated constant flow channel. It can monitor residual chlorine, chlorine dioxide, and ozone in water, using the constant voltage principle for stable and reliable measurement, and easy installation. Mainly used for drinking water, hospital wastewater, secondary water supply, industrial process water, etc.

Product application:Mainly used for drinking water, hospital wastewater, secondary water supply, industrial process water, etc.

Product Features:

No need for pre-treatment of water samples;

No need for any colorimetric reagents;

No need to replace the membrane and electrolyte;

Anti interference dual platinum electrode, reliable measurement;

Constant flow channel maintains stable flow rate;

One machine for three purposes, capable of measuring residual chlorine, chlorine dioxide, and ozone;

● Chinese and English interface, easy to operate;

Password protection to prevent human error;

● 4-20 mA.RS485Modbus;

● One multi-functional relay output;

● Multiple installations including wall mounted, panel mounted, and pipe mounted.

Technical parameters:

measurement range

0.00~20.00ppm,0.00-200.00ppb

resolution

0.001ppm,0.01ppm

precision

±2%F.S

temperature compensation

-10~130 ℃ manual/automatic; (NTC10K/PT1000)

temperature

Working temperature: -10~70.0 ℃; Storage temperature: -20~70.0 ℃

display

Large backlit dot matrix LCD

language

Chinese/English

storage

600000 pieces of data

power supply

90-260VAC,50/60Hz; 24VDC optional

transmitter output

2-channel isolated transmission 4-20mA output, maximum loop 500 Ω, 0.1% F.S, adjustable measurement value and temperature

communication function

RS485 Modbus

Cleaning output

Cleaning interval: adjustable from 0.1-1000h, cleaning time: adjustable from 1-1000s

alarm output

2 independent Hi/Lo alarm points with hysteresis settings, 5A/250VAC/30VDC

Installation method

Wall mounted, pipe mounted, panel mounted

Protection level

IP65

Overall Dimensions

144×144×108mm

Opening size

138×138mm

weight

0.87KG