- Phone
-
Address
5th Floor, Building 5, 599 Huiwang East Road, Jiading District, Shanghai
Gopu Instrument (Shanghai) Co., Ltd
5th Floor, Building 5, 599 Huiwang East Road, Jiading District, Shanghai
Boiler water quality hardness analyzer
Model: Aqualysis800

Boiler water quality hardness analyzer
summary:transportUsing the principle of titration colorimetry and different types of reagents, online monitoring of water hardness and quality control of water softening process are carried out. It has a compact structure, simple operation, high accuracy, low reagent consumption, low maintenance cost, and can operate continuously for a long time. Mainly used for monitoring and controlling the hardness of water quality such as boiler water, softener effluent, cooling circulating water, surface water, and injection water for pharmaceutical factories. Paired with MCS series channel distributors, it can easily achieve hardness measurement of 2-8 channels of water
Product application;Used for monitoring and controlling the hardness of water quality such as boiler water, softener water, cooling circulating water, surface water, and injection water for pharmaceutical factories. Paired with MCS series channel distributors, it can easily achieve 2-8 channel water hardness measurements.
Product Features:
●High cost-effectiveness and high measurement accuracy;
● Automatic calibration, automatic diagnosis, and monitoring;
● Compact design, 400 × 300 × 180mm/4Kg;
Low maintenance and reagent consumption;
Continuous measurement or interval measurement (1-360min);
● 2-channel programmable digital input (can be linked with external devices);
● 4-channel programmable contact output;
● Easy installation and simple operation;
Data storage, fault recording, RS485, 4-20 mA output.
Product Technical Parameters:
Specification Model |
Aqualysis800 |
measurement method |
Titration colorimetric method |
measure time |
2-8 minutes, depending on the hardness of the water |
precision |
± 2% of the upper limit of the selected reagent |
repetitiveness |
± 2% of the upper limit of the selected reagent |
power supply |
85-265VAC,47-63Hz |
power consumption |
25VA (operating), 3.5VA (standby) |
Protection level |
IP65 |
size |
400x300x180mm(HxWxD) |
display |
LCD color screen display |
selectable unit |
°dH、°f、ppmCaCO3、mmol/l |
output |
1. 4-channel programmable relay output (max.250V, 4A) |
2. 1 channel 4-20mA signal, max 750 Ω | |
3. 1 RS485 output | |
input |
1. IN1 input (start analysis) |
2. IN2 input (stop analysis) | |
Analysis Cycle |
Time interval measurement (1-360min)/external signal/flow signal |
Rinse time |
Can be set (15~1800s) |
water quality requirements |
pH:4 – 10 |
Iron:<3 ppm | |
Copper:<0.2 ppm | |
Aluminum:<0.1 ppm | |
Manganese:<0.2 ppm | |
temperature |
Environment: 10 ° C -40 ° C |
Water sample: 5 ° C-40 ° C | |
pressure |
Ca. 0.2-5 bar (max) (recommended 1-2 bar) |