Welcome Customer !

Membership

Help

Zhuhai Hengzhi Electronic Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Zhuhai Hengzhi Electronic Technology Co., Ltd

  • E-mail

    obt002@foxmail.com

  • Phone

    15913270499

  • Address

    Room 502, Taiwan Business Association Center, No. 2029 Mingzhu South Road, Zhuhai City

Contact Now

Natural gas humidity analyzer

NegotiableUpdate on 12/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The OptiPEAK TDL600, a new generation of natural gas humidity analyzer, is specifically designed for fully automatic online humidity measurement of natural gas and biogas, among other biogas.
Product Details

Natural gas humidity analyzerOptiPEAK TDL600 is specifically designed for fully automatic online humidity measurement of natural gas, biogas, and other biogas. The OptiPEAK TDL600 laser analyzer adopts excellent laser spectral absorption principle and signal processing technology, and is a highly stable and high-performance analyzer mainly used for measuring humidity in natural gas. By using non-contact measurement technology, with minimal maintenance costs, even under harsh operating conditions such as constantly changing methane concentrations or the use of an acidic gas, measurement tasks can be completed with ease. This analyzer has obtained certification for all hazardous environments, with high performance, stability, and sensitivity.


feature

D-MET system: responding to changes in gas composition, gas source mixing,

Mixing biogas and shale gas

• Accuracy up to 1 ppmV

• Acid resistant gas environment

• Easy installation and setup features

Low cost maintenance

EExd explosion-proof enclosure, compliant with IECEx, ATEX, and cQPSus certifications

• Complete sampling system

Quality verification by 40 years of humidity measurement experts at Michell


application

• Drying of natural gas

• Natural gas transmission monitoring

• Trade handover

• Natural gas storage

Natural gas offshore transmission pipeline

LNG receiving/re vaporization device

NGL vaporization

• Biogas


Natural gas humidity analyzerOptiPEAK TDL600 Technical Parameters

Measurement Technology TDLAS

Local display 4.3 "color LCD touch screen

Range 1~1000 ppm V

Electrical interface 3 x M20

Accuracy ± 1% reading>100 ppmV verification

± 1 ppmV<100 ppmV factory mode with 3 points traceable to NPL and NIST

Repeatability<1 ppmV (long-term stability<0.1 ppmV/year)

Recommend verification based on user needs to measure a limit of 1 ppmV

Measurement units: ppmV, lb/MMSCF, mg/Nm3, dew point ° C or mechanical index ° F (ISO18453 or IGT # 8)

Sample gas flow rate of 1 Nl/minute, response speed of 1 to 5 Nl/minute, optical feedback of 0.2 seconds, sample gas filtration bypass refresh display of 2~3 seconds

Maximum intake pressure 1450 PSI (100 barg) operating temperature

Indoor version:+10 ° C~+45 ° C

Outdoor version *: -20 ° C~+45 ° C outlet pressure 0.7 to 1.4 bara, filter bypass

Outdoor version with cooling options *: -20 ° C~+55 ° C up to 3 barg maximum

Electrical parameter shell standard/packaging aluminum alloy, explosion-proof shell, IP66, NEMA 4

Gas port 1/4 "NPT (F) voltage 110 V AC or 230 V AC 50/60Hz

Weight 40kg (88lbs) (without sampling system) powered

Indoor installation: 80W sampling system enclosure made of 304L or 316L stainless steel

Outdoor installation: 180W hazardous area certification

Analog signal ATEX: II 2 G Ex d ib op is IIC T5 Gb

Input: 2 x 4-20 mA User configurable IECEx: Ex d ib op is IIC T5 Gb

Output: 3 x 4-20 mA, 3 alarms 250 V AC, 10A Tamb -20 ° C~+60 ° C

cQPSus: Class I, Division 1, Groups A, B, C,

Digital Communication RS485 ModBus RTU D, T5,

TC TR Ex: Tamb -20 ° C~+60 ° C, IP66 data recording. Users can choose the sampling period from 10 seconds to 1 day, recording 1 Ex d ib op is IIC T5 Gb