Dew Point Verification System - Michell DCSIt is a complete calibration system installed on the rack, generating corresponding calibration airflow (or nitrogen) within a pre-set dew point temperature range from a minimum of -100 ° C (-148 ° F) to a maximum of+20 ° C (+68 ° F). When ordering this system, customers only need to specify the required operating range, and MIZIL will determine an economical way to create your verification system.
As a reliable standard, the standard hygrometer provides calibration gas for dew point generators for measurement and can be compared with the instrument being tested. The cold mirror hygrometer directly measures the temperature generated by condensate and can provide repeated and reliable measurement results every time. This means that this instrument is suitable for use as a standard instrument. To ensure traceability to higher standards, traceable national standards or UKAS recognized standards can be provided.
DCS60 is equipped with S8000 Integral, which uses air cooling for heat dissipation and can achieve dew point measurement at -60 ° C (-76 ° F).
DCS80 is equipped with S8000 RS, which features an automatic control auxiliary cooling system and can achieve dew point measurement of -90 ° C (-130 ° F).
DCS100 is equipped with S4000 TRS, a semi-automatic control auxiliary cooling system, and dew point measurement at -100 ° C (-148 ° F).
highlight
A complete dew point calibration scheme can be selected from compressors, dryers, dew point generators, standard instruments, and optional multi way valves
Fast response time to changes in the set point output
• Reliable humidity generator
• Easy to operate through manual mixing of flow or setting point buttons
Remote control via RS232 communication (depending on model)
Dew Point Verification System - Michell DCSparameter
DCS60 |
DCS80 |
DCS100 |
|
range |
-60 ~ +20°C dp |
-80 ~ +20°C dp |
-100 ~ +20°C dp |
Air dryer |
PSD2 dryer |
PSD4 Super Drying Machine |
|
Mode of occurrence |
DG2 manual flowmeter |
VDS system with flow control |
|
Standard humidity meter |
S8000 cold mirror hygrometer |
S8000 RS cold mirror hygrometer |
S4000 TRS cold mirror hygrometer |
Verify traffic |
2 ~ 5 Nl/min |
10 Nl/min |
|
Optimize system uncertainty |
± 0.2 ° C (± 0.36 ° F) Dew point (k=2) @+20 ° Cdp (+68 ° Fdp) |
||
Set point accuracy |
Dew point at ± 0.5 ° C (± 0.9 ° F) |
||
carrier gas |
Oil free compressed air (supplied by compressor) |
||
Operating Temperature |
+15 ~ +30°C (+59 ~ +86°F) |
||
trace to the source |
Traceable to NPL and NIST through standard hygrometers |
||
power supply |
220~240 V AC or 100~130 V DC, 50/60 Hz |
||
shell |
19 "rack with wheels, 1.9 m (74.8") high |
||
weight |
98kg (216lbs) (approx) |
125kg (231lbs) (approx) |
undecided |