- Phone
-
Address
Baoding National High tech Industrial Development Zone
Baoding Lange Constant Flow Pump Co., Ltd
Baoding National High tech Industrial Development Zone
LSP12-XBMulti channel digital laboratory injection pumpCharacteristics
Color LCD screen display, full touch screen control.
Small size, supports vertical screen operation, saves experimental table space
High driving force (290N) and adjustable thrust (20% -100%)
Wide flow range, real-time flow adjustment can be performed during pump operation
Stable flow with small fluctuations, capable of high-precision transmission of trace liquids
Provide LabVIEW drivers and demos to facilitate the construction of LabVIEW control and monitoring systems
Strong electromagnetic compatibility performance, suitable for strong electromagnetic radiation environments such as electrospinning
Support external signal control and communication control (RS485 interface, Modbus RTU)、 Foot switch control
Intuitively indicate the working status of the pump through buzzing sound, lighting, text information, etc
Built in multiple common brand syringes, supports custom syringe parameter saving, and improves flow accuracy by calibrating custom syringes
Formula parameters can be saved as schemes for easy access
The software of the pump can be upgraded on-site through a USB flash drive
LSP12-XBMulti channel digital laboratory injection pumpDetailed specifications
Product Model |
LSP12-XB |
|||
Working Mode |
Single injection, multiple uniform injections, single extraction, injection before extraction (cycle), extraction before injection (cycle) |
|||
Number of syringes |
4 (Four channel clamping accessories) |
6 (Six channel clamping accessories) |
8 (Eight channel clamping accessories) |
10 (Ten channel clamping accessories) |
Applicable syringe specifications |
10μL-10mL |
|||
Reference flow range |
69pL/min-39mL/min |
|||
maximum stroke |
140mm |
|||
Travel resolution |
0.0326 μ m/microstep |
|||
Linear velocity range |
0.416μm/min-180mm/min |
|||
Linear velocity adjustment resolution |
0.416μm/min |
|||
Accuracy and precision of travel control |
Error ≤± 0.35% (travel ≥ 30% of maximum travel) |
|||
Repetitive accuracy of travel control |
CV ≤ 0.03% (travel ≥ 30% of maximum travel) |
|||
Maximum linear thrust |
290N (thrust can be set to 20% -100%) |
|||
Syringe selection |
Selection of main manufacturers and models of built-in syringes |
|||
Syringe customization |
Can save 10 sets of custom syringes, syringe specifications: stroke ≤ 140mm, 6.2mm ≤ outer diameter ≤ 20mm |
|||
flow calibration |
Calibrate custom syringes to improve flow and liquid volume accuracy |
|||
display mode |
4.3-inch color LCD displays parameters such as flow rate, liquid volume, and time |
|||
Parameter scheme saving |
Can save 10 sets of parameter schemes |
|||
Control method |
Touch screen control, external signal control, communication control, foot switch control, Labview control |
|||
External control function |
Control: Two control inputs, relay signals or level signals control start stop and emergency stop Status output: Two status output signals that can be configured as start stop, direction, or fault status at will |
|||
communication control |
RS485 interface, Modbus RTU or Lange OEM protocol |
|||
Real time traffic regulation |
Support real-time traffic regulation during operation |
|||
Sound prompt function |
Multiple sound prompt functions can be set: button sound, running completed, about to complete, traffic jam, equipment failure |
|||
status indicator light |
Green - Running, Blue - Standby, Red - Fault |
|||
Lock screen function |
Prevent erroneous operations |
|||
Vertical screen function |
Support vertical placement for use, with a 90 degree rotation display on the screen interface |
|||
power-on state |
Power on stop |
|||
Fast forward, fast rewind |
Full speed injection or full speed extraction |
|||
Electromagnetic Compatibility |
Certified by NB CE/UKCA |
|||
Overall dimensions |
280mm×223mm×160mm |
280mm×263mm×160mm |
280mm×270mm×160mm |
280mm×315mm×160mm |
Adapter power supply |
AC 90V-260V |
|||
Pump power supply |
DC24V/15W |
|||
Working environment temperature |
0 - 40°C |
|||
Working environment humidity |
≤80%RH, no condensation |
|||
weight |
4.65kg |
4.85kg |
4.95kg |
5.15kg |
Applicable syringe specifications
Applicable syringe specifications |
Inner diameter of syringe (mm) |
Reference flow range |
10μL |
0.46 |
0.0691nL/min-0.0353mL/min |
25μL |
0.80 |
0.2090nL/min-0.0904mL/min |
50μL |
1.10 |
0.3951nL/min-0.1710mL/min |
100μL |
1.60 |
0.8360nL/min-0.3617mL/min |
250μL |
2.30 |
1.7275nL/min-0.7475mL/min |
500μL |
3.25 |
3.4493nL/min-1.4925mL/min |
1mL |
4.72 |
7.2752nL/min-3.1479mL/min |
5mL |
13.10 |
56.041nL/min-24.248mL/min |
10mL |
16.60 |
89.987nL/min-38.937mL/min |