1. All functions can be directly accessed
2. A large number of interchangeable probes
3. High contrast display screen and backlit keyboard
4. Storage and statistical functions
5. Bluetooth and USB functionality
6. Durable casing with IP 65 protection level

Measurable according to probe model:
▪ Non magnetic coatings on steel and insulation coatings on non-ferrous metals.
▪ Thin coatings of non-ferrous metals, oxides, or paint on small steel workpieces. (F 05)
▪ Non magnetic coatings, thick coatings, and enamel coatings on steel.
▪ A precise solution for very thin coatings. The insulation layer on non-ferrous metals, such as paint film, enamel or anodized layer. (N 02)
▪ A thick plastic like coating on tanks, pipelines, and container buildings. (F 10)
▪ Thick plastic, rubber or concrete layers and corrosion-resistant layers in pipeline construction. (F 20)
▪ Very thick corrosion-resistant layer and anti bulging layer. (F 50)
▪ Special high-temperature probes allow for coating thickness measurement on hot surfaces with surface temperatures up to 250 ° C or 350 ° C (F 2 HT)
| MiniTest 2500/MiniTest 4500 Host Technical Specifications | ||
|
|
MiniTest 2500 |
minitest 4500 |
|
data storage
-Total number of storable readings
-Large number of groups
-Having independent calibration values
Application storage quantity
-Batch groups with the same calibration values
The number of batch groups stored in each application
|
2.000.000
1
-
-
|
2.000.000
Over 9500
99
99
|
|
Statistical function (per batch group) |
kvar, n, max., min.
kvar, n, max., min.
|
each, n, max., min., CP, CPK
each, n, max., min., CP, CPK
|
|
calibration |
Factory calibration, zero point and up to four calibration points |
If there is no substrate to use, calibrate through coating
(CTC)
|
|
Compensation function |
- |
Used to add/subtract a constant value from a reading |
|
Extreme settings (user-defined)
Having monitoring function
|
- |
Alarm for light and sound signals exceeding the time limit |
|
unit of measurement |
μm, mm, cm, mils, inch |
|
|
Interface |
USB |
USB and Bluetooth 4.0 |
|
Upgradeable interface |
- |
Alarm output, foot switch triggered, RS 232 interface |
|
Power supply |
3 x AA (LR06) batteries, the working time for USB battery replacement is approximately 150 hours (without backlight) |
|
|
Norms and standards |
DIN EN ISO 1461, 2064, 2178, 2360, 2808, 3882; ISO 19840; ASTM B 244,
B 499, D 7091, E376
|
|
|
display |
53 x 46 mm, backlight |
|
|
Operating temperature/storage temperature |
–10 °C … 60 °C / –20 °C … 70 °C |
|
|
Size/Weight |
153 mm x 89 mm x 36 mm/320 g (instrument including battery), rubber protective cover 90 g |
|
|
Protection level |
IP 65 |
|
| F-type probe (magnetic induction principle) |
||||||
| probe | Measurement range | Low end resolution: | Ensure tolerance (reading): | *小曲率半径(凸/凹) | *Small measurement area | *Small substrate thickness |
| F05 | 0...500 μm | 0,1 μm | ± (1% of reading + 0,7 μm) | 0,75 mm / 5 mm | Ø 3 mm | 0,1 mm |
| F1.6 | 0...1600 μm | 0,1 μm | ± (1% of reading + 1 μm) | 1,5 mm / 10 mm | Ø 5 mm | 0,5 mm |
| F3 | 0...3000 μm | 0,2 μm | ± (1% of reading + 1 μm) | 1,5 mm / 10 mm | Ø 5 mm | 0,5 mm |
| F1.6/90 | 0...1600 μm | 0,1 μm | ± (1% of reading + 1 μm) | Flat/6mm | Ø 5 mm | 0,5 mm |
| F1.6P | 0...1600 μm | 0,1 μm | ± (% 3% v. messwert + 1 μ m) | Only on a flat surface | Ø 30 mm | F 0,5 mm N 0,5 mm |
| F2/90 | 0...2000 μm | 0,2 μm | ± (1% of reading + 1 μm) | Flat/6mm | Ø 5 mm | 0,5 mm |
| F10 | 0...10 mm | 5 μm | ± (1% of reading + 10 μm) | 5 mm / 16 mm | Ø 20 mm | 1 mm |
| F20 | 0...20 mm | 10 μm | ± (1% of reading + 10 μm) | 10 mm / 30 mm | Ø 40 mm | 2 mm |
| F50 | 0...50 mm | 10 μm | ± (3% of reading + 50 μm) | 50 mm / 200 mm | Ø 300 mm | 2 mm |
| N-type probe (eddy current principle) | ||||||
| probe | measurement range | Low end resolution | Ensure tolerance (reading): | *小曲率半径(凸/凹) | *Small measurement area | *Small substrate thickness |
| N.08 Cr | 0...80μm | 0,1μm | ± (1% of reading + 1 μm) | 2,5 mm / 10 mm | Ø5 mm | ≥100 μm for Cu on Fe |
| N 02 | 0...200 μm | 0,1 μm | ± (1% of reading + 0,5 μm) | 1 mm / 5 mm | Ø 0,2 mm | 50 μm |
| N 1.6 | 0...1600 μm | 0,1 μm | ± (1% of reading + 1 μm) | 1,5 mm / 10 mm | Ø 5 mm | 50 μm |
| N1.6/90 | 0...1600 μm | 0,1 μm | ± (1% of reading + 1 μm) | flat / 6 mm | Ø 5 mm | 50 μm |
| N2/90 | 0...2000 μm | 0,2 μm | ±( 1% of reading + 1 μm) | flat / 6 mm | Ø 5 mm | 50 μm |
| N10 | 0...10mm | 10 μm | ± (1% of reading + 25 μm) | 25 mm / 100 mm | Ø 50 mm | 50 μm |
| N20 | 0...20 mm | 10 μm | ± (1% of reading + 50 μm) | 25 mm / 100 mm | Ø70 mm | 50 μmm |
| N100 | 0...100 mm | 100 μm | ± (3% of reading + 0,3 mm) | 100 mm / flat | Ø 200 mm | 50 μm |
| CN02 | 10...200 μm | 0,2 μm | ± (3% of reading + 1 μm) | Only on a flat surface | Ø 7 mm | none |
| FN probe (dual function type) | ||||||
| probe | measurement range | Low end resolution | Ensure tolerance (reading): | *小曲率半径(凸/凹) | *Small measurement area | *Small substrate thickness |
| FN1.6 | 0...1600μm | 0,1 μm | ± (1% of reading + 1 μm) | 1,5 mm / 10 mm | Ø 5 mm | F 0,5 mm N 50 μm |
| FN1.6P | 0...1600 μm | 0,1 μm | ± (3% of reading + 1 μm) | nur flate Flachen | Ø 30 mm | F 0,5 mm N 0,5 mm |
| FN 1.6/90 | 0...1600 μm | 0,1 μm | ± (1% of reading + 1 μm) | flat / 6 mm | Ø 5 mm | F 0,5 mm N 50 μm |
| FN2/90 | 0...2000 μm | 0,2 μm | ±(1% of reading + 1 μm) | flat / 6 mm | Ø 5 mm | 0,5 mm |
▪ Non magnetic coatings on steel and insulation coatings on non-ferrous metals.
▪ Thin coatings of non-ferrous metals, oxides, or paint on small steel workpieces. (F 05)
▪ Non magnetic coatings, thick coatings, and enamel coatings on steel.
▪ A precise solution for very thin coatings. The insulation layer on non-ferrous metals, such as paint film, enamel or anodized layer. (N 02)
▪ A thick plastic like coating on tanks, pipelines, and container buildings. (F 10)
▪ Thick plastic, rubber or concrete layers and corrosion-resistant layers in pipeline construction. (F 20)
▪ Very thick corrosion-resistant layer and anti bulging layer. (F 50) ▪ Special high-temperature probes allow for coating thickness measurement on hot surfaces with surface temperatures up to 250 ° C or 350 ° C. (F 2 HT)