Microcomputer measurement and control instrument, gas and electricity meterWe mainly provide customized non-standard measuring tools based on customer needs. For example, non-standard ring gauges, plug gauges, pneumatic gauges, spline gauges, irregular gauges, thread depth touch gauges, taper gauges, etc. Design and produce according to requirements.



Environment: Temperature20±1℃Humidity40%–60%No vibration, no dust, no oil stains.
Gas source: pressure0.4–0.6 MPaConstant, throughDrying+filteringNo oil or water to prevent nozzle blockage.
Instrument: placed inSeismic platform, probe and workpiece surfaceverticalThe workpiece is clean and free of burrs and oil stains.
Probe: screw into the installation hole to ensureNo looseness or air leakage.
Power on: Turn on the power and wait3-5 minutesPreheating and system initialization.
Parameters: Enter the menu, setMeasurement unit, range, resolution, upper and lower limits of tolerance.
Zero setting: Place the probe inLower limit standard componentsPress' zero/lower limit 'to adjustZero position knobAlign the displayed value with the standard value and lock it.
School upper limit: probe insertionUpper limit standard partsAdjust according to the 'school upper limit'Magnification knobMatch the displayed value with the standard value.
Review: Repeat the insertion of upper and lower limit standard parts untilBoth points are accurateSave calibration.
Only insert the lower limit standard parts and press' zero 'to reset them;Replace the probe/gas source/environmental changesTwo point calibration must be performed.
Positioning: Probeslowly approachWorkpiece, maintainverticalAvoid colliding with the nozzle.
Reading: To be displayedstableRead the numerical value afterwards; Can record or export data.
Batch: When measuring continuously,Every 10-20 piecesVerify the zero position once using standard parts.
Judgment: According to the preset tolerance,Green=qualified, red=out of tolerance.
Shutdown: First turn off the gas source, then turn off the instrument power.
Cleaning: Gently blow the nozzle of the measuring head with compressed air to clean the instrument and workbench.
Storage: probe cover protective cover, instrument cover dust cover.
Unstable data: checkGas source pressure, leakage, probe perpendicularity, workpiece cleanliness.
Indication deviation: redoTwo-point calibrationCheck if the standard parts are worn.
Nozzle blockage: Use a fine needle to gently pass through,Do not stab with hard objects.
Strictly prohibited: probeImpact workpieceGas source with oil and water, severe environmental vibration.