- Phone
-
Address
No. 1210 Dongfang Avenue, Wuzhong Economic Development Zone, Suzhou
Suzhou Changliu Purification Technology Co., Ltd
No. 1210 Dongfang Avenue, Wuzhong Economic Development Zone, Suzhou
Suzhou Changliu Purification Floating Bacteria Sampler Manufacturer JYQ-IVMicrobial Air Monitoring SystemsCompliant with the GMP Good Manufacturing Practice for Pharmaceutical and Food Products - ISO 14698-1/2 Control of Biological Pollution in Cleanrooms and Related Control Environments - GB/T 16293-2010 Testing Requirements for Floating Bacteria in Cleanrooms (Areas) of the Pharmaceutical Industry.

JYQ-IVMicrobial Air Monitoring SystemsAn efficient porous inhalation dust and bacteria sampler. It is designed based on the principle of particle impact and the theory of constant velocity sampling. Sampling is direct, and the wind speed at the sampling port is basically the same as that in the clean room, which can more accurately reflect the microbial concentration in the clean room. During sampling, the air carrying dust bacteria passes through the micropores at high speed and is impacted on the surface of the agar in the culture dish; These live microorganisms undergo dynamic rehydration during the cultivation process, growing at high speeds to obtain faster results.
The JYQ-IV type air bacterial sampler has a novel structure and is divided into two parts: an upper part with a collection port, sampling seat, and air pump, and a lower part with a controller and battery. The sampling port and housing are made of high-quality aviation aluminum, with a closed cell surface treatment for easy sterilization and disinfection before use. This instrument has the functions of *, large sampling volume, stable performance, easy operation, and reaches the advanced level of international similar products. It is an ideal floating dust bacteria concentration sampling instrument for various pharmaceutical factories, hospitals, biological products, food processing, public places and other testing departments.
JYQ-IV type meets the standard
Good Manufacturing Practice (GMP) for Drug Production Quality Management
ISO14698-1/2 Control of Biological Pollution in Cleanrooms and Related Control Environments
GB/T 16293-2010 Test Method for Floating Bacteria in Cleanrooms (Areas) of the Pharmaceutical Industry


Technical characteristics of JYQ-IV type:
1. Original German EBM imported fan, electronic flow sensor, high flow control accuracy;
2. Programmable, sample size can be set arbitrarily from 0-9999 L
3. Real time temperature and pressure detection, and automatic compensation of sampling flow rate; Automatically stop sampling when the rated sampling flow rate cannot be reached (due to insufficient power or blocked sampling head) to ensure sampling accuracy;
4. Dual use AC/DC, rechargeable battery, can work continuously for 4-6 hours after being fully charged
5. LCD display sampling quantity, sampling time and other parameters
6. Power saving function, automatic adjustment of backlight brightness, automatic shutdown without operation;
7. All metal shell, solid and durable, stylish and beautiful;
8. The sampling head can be adjusted at any angle from 0 ° to 90 °;
9. Store parameters such as sampling volume and sampling time in groups, up to 256 sets of data can be stored
Main parameters |
parameter range |
resolution |
accuracy |
sampling flow |
100L/min |
1L/min |
Better than ± 2.5% |
Environmental atmospheric pressure |
(80~106)kPa |
||
Operating Temperature |
(-20~50)℃ |
||
Specification of culture dish |
Ф90mm |
||
material |
aviation aluminum |
||
lithium battery pack |
16.8V 5200m A |
||
Charging time |
<4h |
||
Battery working time |
4-6h |
||
External power supply |
16.8V2A |
||
Host size W × D × H |
(Ф120×300)mm |
||
instrument noise |
<60dB(A) |
||
Total weight |
About 3kg |
||
power consumption |
20W |
||