The medical RK7305Y grounding resistance tester is used to measure the grounding resistance inside electrical equipment. It reflects the (contact) resistance between the exposed conductive parts of the electrical equipment and the main grounding terminal of the electrical equipment. This product meets safety standards such as GB 4706.1-2005 and GB 9706.1-2007, and complies with the requirements of JJG 984-2004 Safety Measurement and Verification Regulations.
Product Introduction
RK7305Y MedicalGrounding resistance testerIt is used to measure the grounding resistance inside electrical equipment, which reflects the (contact) resistance between the exposed conductive parts of the electrical equipment and the total grounding terminal of the electrical equipment.
This product meets safety standards such as GB 4706.1-2005 and GB 9706.1-2007, and complies with the requirements of JJG 984-2004 Safety Measurement and Verification Regulations.
Application field
1household appliancesTelevision, refrigerator, air conditioner, washing machine, dehumidifier, electric blanket, charger, etc
2transformer:Oscilloscope, signal generator, DC power supply, switch power supply and other types of complete machines
3Medical equipment:Various new medical instruments and medical supporting instruments, cardiac monitoring, medical imaging, biochemical analysis instruments, thermometers and other home medical equipment
4Lighting industry:Various lighting fixtures such as ballasts, road lights, stage lights, handheld lights, etc
5New energy vehicles:Electric vehicle battery pack connection bridge, cell connection resistance
6Electronic components:Diodes, transistors, high-voltage silicon stacks, various electronic transformers, connectors, high-voltage electrical appliances
7Electric heating appliances:Electric drills, cutting machines, grinding machines, welding machines, etc
RK7305Y MedicalGrounding resistance tester
Input power supply: 50/60Hz ± 5% 115/230VAC ± 10%
Output current: 3-30Aac constant current source, resolution: 0.1A/step, accuracy: ± (2% set value+0.02A)
Output voltage: 6Vac MAX (measured when open circuit)
model |
RK7305Y |
input power |
50/60Hz±5% 115/230VAC±10% |
output current |
3-32Aac constant current source, resolution: 0.1A/step, accuracy: ± (2% set value+0.02A) |
output voltage |
6Vac MAX (measured when open circuit) |
output frequency |
50Hz/60Hz selectable |
Resistance range |
0-600m Ω, output current 3-10A; 0-120m Ω, output current 10A-32A, resolution: 1m Ω/step
|
Zero point adjustment |
MAX 100m Ω, resolution: 1m Ω/step, accuracy: ± (2% reading value+1 word) |
Upper limit setting |
Range: 0-600m Ω, Resolution: 1m Ω/step, Accuracy: ± (2% set value+1 word) |
Test time |
Scope:0.5-999.9 seconds (0=continuous test) |
control interface |
Input: Test(TEST)、 Resetting and Withstanding Voltage Test |
|
PROCESSING)。 Output: Test passed (PASS), test failed (FAIL)
Testing in Progress (TEST IN ROCESS)
|
Test result output method |
BuzzerLCD displays numbers and outputs control interface status |
Memory group |
5 sets of test condition memory |
Correction method |
Software calibration |
monitor |
16 × 2 backlit LCD display |
temperature and humidity |
0℃~40℃,≤75%RH |
Size(D×W×H) |
370mm×290mm×90mm |
Standard Appendix |
RK00001 power line, RK00005 test line, RK00002 cable |
Safety Lock |
Equipped with keyboard lock function |
weight |
8.85KG |
fuse |