- Phone
-
Address
Room A609, Puchen Building, No. 12 Kaihua Road, Huayuan Industrial Zone, Tianjin
Tianjin Nikos Testing Technology Co., Ltd
Room A609, Puchen Building, No. 12 Kaihua Road, Huayuan Industrial Zone, Tianjin
BRT-300Mask respiratory resistance testerstandard GB 2626-2019:6.5+6.6、GB/T 32610-2016:6.7+6.8、EN 149:8.9、AS/NZS 1716:2012- 3.2.5、GB/T 16556-2007:6.13、GB 30864-2014:6.5。
Specifications
1. Compressed air regulation range (0.1-0.6) Mpa;
2. TSI gas flowmeter from the United States: flow rate (0-300) L/min output adjustable, accuracy: 1.5%. Can complete respiratory resistance tests at different flow rates of 30, 85, 95, 160, and 300 L/min;
3. Differential pressure gauge produced in the United States: ± 750pa, accuracy: 0.25% F.S;
4. Silicone head mold (S, M, L sizes optional, including breathing tube);
5. Equipped with a negative pressure suction device, the suction flow rate can be adjusted within a range of (10~120) L/min;
6. Size (length * width * height cm): 60 * 38 * 70;
7. Weight (kg): 80.
Installation requirements:
1. Power supply: AC 220V 50Hz 500W
2. Environment: 5-35 ℃
3. Compressed air
BRT-300Mask respiratory resistance testerconfiguration
1. 1 host
2. Standard configuration: 1 M-size head mold
3. 1 copy of product manual
The following calibration related items meet the standard requirements:
Exhalation resistance (gauge pressure) (Pa)
Inhalation resistance (gauge pressure) (Pa)
Exhalation flow rate (L/min)
Inhalation flow rate (L/min)
airtightness
Appearance requirements