The RK2675 series leakage current tester is used to measure the leakage current generated by the working power supply (or other power sources) of electrical appliances through insulation or distributed parameter impedance, which is unrelated to work. It is a testing experimental equipment used to measure the insulation performance and product safety performance of electrical appliances. This product meets domestic and international safety standards such as GB4706.1-2005 and IEC60335.1-2004, and complies with the requirements of JJG843-2022 Leakage Current Metrology Verification Regulations and Safety Metrology Verification Regulations.
Product Introduction
RK2675 seriesLeakage current testerThe leakage current generated by the working power source (or other power sources) of electrical appliances through insulation or distributed parameter impedance, which is unrelated to work, is a testing experimental equipment used to measure the insulation performance and product safety performance of electrical appliances.
This product meets domestic and international safety standards such as GB4706.1-2005 and IEC60335.1-2004, and complies with the requirements of JJG843-2022 Leakage Current Metrology Verification Regulations and Safety Metrology Verification Regulations.
Application field
Household appliances: TV, refrigerator, air conditioner, washing machine, dehumidifier, electric blanket, charger, etc
Instruments and meters: oscilloscopes, signal generators, DC power supplies, switching power supplies, and other types of complete machines
Lighting appliances: ballasts, road lights, stage lights, handheld lights, and various other lighting fixtures
Electric heating appliances: electric drills, gas cutting machines, polishing machines, grinders, welding machines, etc
Motor: Rotating motor, etc
Performance Characteristics
1. Can measure the leakage current of various low-voltage electrical equipment
2. The test voltage, leakage current, and test time are all displayed digitally
3. Leakage current value and testing time can be pre-set
4. Leakage current exceeding the limit with sound and light alarm
5. Test phase can automatically/manually switch
6. The testing time and alarm value can be continuously set
RK2675AM/RK2675WM/RK2675B/RK2675C/RK2675D/RK2675ELeakage current testerOutput voltage: 0-250V
Test current: 0.03-2mA/20mA
Test accuracy: ± 5%
Product Description Technical Parameters RK2675AM/B/C/D/E Operation Video RK2675W
model |
RK2675WM |
RK2675AM |
RK2675B |
RK2675C |
RK2675D |
RK2675E |
output voltage |
0~300V |
0~250V |
test current |
0.03~2mA/20mA |
test accuracy |
±5% |
Test time |
0.0s~999s0.0=continuous test |
1~99s±1% |
transformer capacity |
passive
(Maximum external 6000VA isolated power supply)
|
500VA |
1000VA |
2000VA |
3000VA |
5000VA |
PLC interface |
optional |
none |
Power requirements |
220V±10% 50Hz±5% |
working environment |
0℃~40℃,≤85%RH |
Overall dimensions(DxWxH) |
318*248*176mm |
316*253*166mm |
410*377*200mm |
493*436*240mm |
500*440*236mm |
514*469*249mm |
weight |
4.95Kg |
10Kg |
22.8Kg |
39.1Kg |
45.75Kg |
68.15Kg |
accessories |
power cord |
optional |
Optional selection(PLC interface includes RK00002 connection line, RK401 inspection box, RK00017 (passive leakage testing line) |