E-mail
1276198359@qq.com
-
Phone
13621795675
-
Address
Room 202, Building B, Zhengjue Technology Building, Lane 1356, Xinyuan Road, Minhang District, Shanghai
Shanghai Hansen Scientific Instrument Co., Ltd
1276198359@qq.com
13621795675
Room 202, Building B, Zhengjue Technology Building, Lane 1356, Xinyuan Road, Minhang District, Shanghai
1、 Wide testing range, capable of testing particles between 2~100um. With a small sample size, efficient and accurate particle data can be obtained.
IIKS-42AF can measure hydrofluoric acid liquid particlesApplication parameter analysis:
Can detect particles with a small diameter of 2um
The customizable particle size range is 2~100um, which can be expanded to 10 channels through the controller of KE-40B1
The factory setting is 8-channel: ≥ 2 μ m, ≥ 3 μ m, ≥ 5 μ m, ≥ 7 μ m, ≥ 10 μ m, ≥ 25 μ m, ≥ 50 μ m, ≥ 100 μ m
(≥150μmsupportavailableasoption)
Integrated leakage sensor and alarm output
KS-BF sapphire material capillary has stronger corrosion resistance
IIIKS-42AF can measure hydrofluoric acid liquid particlesProduct Description: How to Translate
| Detection method | Light scattering method |
| light source | Laser diode (wavelength 830nm, output power: 200MW) |
| Laser grade | Class1,IEC 60825-1 |
| Optical detector | PIN type photodiode |
| Material of parts in contact with the sample | KS-4D: Quartz PFA、PTFE
|
| Types of testable samples | Any liquid that does not corrode or come into contact with the material of the sample components |
| correction | PSL standard solution (particle size 1.6) |
| Particle size range | 2~100um |
| 8 channels (factory settings) | ≥2 μm, ≥3 μm, ≥5 μm, ≥7 μm, ≥10 μm, ≥25 μm, ≥50 μm, ≥100 μm
|
| Testeable range | 2~100um (expandable to 10 channels) |
| Detection efficiency | 70%~110% |
| sampling quantity | 10mL/min |
| Testeable particle concentration | 1200 pieces/mL |
| Sample temperature range | 15 ℃~35 ℃ (no solidification or condensation inside the capillary tube) |
| Sample testing pressure range | Not exceeding 300kPa (gauge pressure) |
| Sample in and out | 2 (diameter) * 4 (diameter) pipeline |
| Gas cleaning device | Rc1/8 ( 1/8PT female screw) |
| Data transmission and operation interface | Transferred to KE-40B1, direct touch screen operation without the need for an external computer |
| Data output method | Built in printer/memory card storage |
| Leakage alarm conversion | Alarm port short circuit during normal testing, open to detect internal leakage |
| power supply | DC 12V (provided by KE-40B1) |
| Test environment conditions | 15-35 ℃, humidity not exceeding 85% (no condensation) |
| Size&Weight | 125 (H) * 140 (W) * 150 (D) mm (excluding protruding parts), About 2.2kg |
| Factory standard configuration | Pipeline accessory combination * 1, signal line (1m) * 1, power line * 1 |
| optional | Signal line B (5m) KS-42-123 |