Welcome Customer !

Membership

Help

Suzhou Changliu Purification Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Suzhou Changliu Purification Technology Co., Ltd

  • E-mail

  • Phone

  • Address

    No. 1210 Dongfang Avenue, Wuzhong Economic Development Zone, Suzhou

Contact Now

Industrial spray laser particle size analyzer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Winner319 industrial spray laser particle sizer is an upgraded product of Winner318 series of spray laser particle sizer designed and developed by our company for droplet particle size measurement. This instrument adopts Mie scattering, Fraunhofer diffraction principle and typical parallel light path design, together with high-performance high-power laser source, to meet the test needs of complex sample environment and large droplet size distribution. Winner319 is a large-scale, open type spray laser particle size analyzer for droplet size measurement.

Product Details

Product Introduction:

Winner319Industrial spray laser particle size analyzerThe spray laser particle sizer designed and developed by our company for the measurement of droplet size is an upgraded product of Winner 318 series. This instrument adopts Mie scattering, Fraunhofer diffraction principle and typical parallel light path design, together with high-performance high-power laser source, to meet the test needs of complex sample environment and large droplet size distribution. Winner319 is a large-scale, open type spray laser particle size analyzer for droplet size measurement.

Scope of application:

Winner319 is suitable for fire protection, forest vegetation, pesticide detection, aeroengine, spray granulation, nozzle research, production of various aerosols, oil mist, water mist, smoke, spray drying and other fields.

Industrial spray laser particle size analyzerProduct features and advantages:

Independently developed Zhuanli technology

By using typical parallel light testing technology and spectrum amplification technology, the range can be greatly expanded in a limited space, and multiple auxiliary integrated photodetectors can be added to effectively collect scattered light from various angles corresponding to the testing range, achieving testing accuracy and reliability throughout the entire range.

*The airflow protection device

Designed with an airflow protection system, it can effectively protect the lens and prevent fog droplets from contaminating the lens during testing.

Split structure design

The split structure and adjustable test area can meet the needs of spray test in any situation, and have the characteristics of non-contact measurement of droplet particles and non-interference with the flow field.

Stable automatic centering system

Reduced the adjustment difficulties caused by the movement of the optical path, and completed the alignment of the optical path with just one click.

Multiple granularity distribution models

You can freely choose the standard mode, RR distribution, and log normal distribution. It can achieve free conversion between volume distribution and quantity distribution.

Technical parameters:

Specification Model

Winner319A

Winner319B

Winner319C


execution standard

ISO13320:2009,GB/T19077-2016,Q/0100JWN001-2013


instrument structure

split-type


test scope

1-500μm

1-1000μm

1-2000μm


Number of detector channels

72

75

80


Accuracy error

≤ 1% (D50 value of national standard sample)


repeatability error

≤ 1% (D50 value of national standard sample)


laser

LD pumped laser; λ=532nm; Power adjustable from 1-40mw


Injection method

open-ended


Length of measurement area

0.1-10m


Lens protection

Double air curtain type






volume

Radiation end: 369mm * 295mm * 360mm

Receiver: 850mm * 280mm * 380mm

weight

45Kg+23Kg