Welcome Customer !

Membership

Help

Zhuhai Hengzhi Electronic Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Zhuhai Hengzhi Electronic Technology Co., Ltd

  • E-mail

    obt002@foxmail.com

  • Phone

    15913270499

  • Address

    Room 502, Taiwan Business Association Center, No. 2029 Mingzhu South Road, Zhuhai City

Contact Now

Hydrogen oxygen analyzer

NegotiableUpdate on 12/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Hydrogen oxygen analyzer
Product Details

Hydrogen oxygen analyzerThe XTP601-EX rugged, linear, and stable XTP601 oxygen analyzer is suitable for measuring gases such as hydrogen, nitrogen, biogas, or carbon dioxide. The sensors of this analyzer are installed in an IP66 rated enclosure, making it suitable for a wider range of fields. After installing the flame arrester, the analyzer has explosion-proof performance and can be used in hazardous environments.


highlight

• Ordinary or ATEX, IECEx, CCSAUS, TC, TR&Ex hazardous area certification

• Optional range: 0-0.5%, 0-50%, 20, 80 or 90-100%

Touch sensitive display that can be calibrated or adjusted without the need for explosion-proof certification application

Excellent accuracy, better than 1% of full scale

• Zero drift stability, with monthly drift less than 0.25%

• Output includes RS485 and two 4-20mA channels (under NAMUR NE 44 standard)

Low maintenance cost, requiring minimal maintenance


application

Monitoring of inert protective gases during hydrocarbon production process

Inert gas monitoring in the pharmaceutical or chemical industry

• Biogas, landfill sites, and biogas plants

Furnace gas control in the metallurgical industry

• Catalytic reforming

N2 generator

O2 generator



Hydrogen oxygen analyzerTechnical Specifications

performance

Measurement technology - Thermomagnetic oxygen sensor

Gas process gas sampling, non condensing, solid particles<5 μ m

Measurement range from 0-0.5% up to 0-50% 20, 80 or 90-100%

Display resolution 0.01% for standard range 0.1% for over range

Display Type Backlight LCD

Accuracy

(excluding special ranges) ± 0.02% oxygen content or<± 1%, whichever is greater

Special range accuracy (80/90-100%)<0.2% O2

Response time (T90) HSR function needs to be enabled within 15 seconds

Repeatability ± 0.2% or 0.01% oxygen content, whichever is greater

Linear ± 0.5% or 0.05% oxygen content, whichever is larger

Zero stability ± 0.25% of monthly range

Range stability ± 0.25% of monthly range

Sample gas flow rate GP1 and GP3: 100! 500 ml/min

EX1, EX3, and GP2: 270~330 ml/min

Sample gas pressure 0.75~2 BarA

Sample gas temperature 5~+55 ° C

Standard maximum (+45 ° C)

Sampling room temperature standard+50 ° C (optional range and gas of+55 and+60 ° C)

The background gas analyzer is calibrated under process background gas and standard gas

For N2, CO2, CH4, H2&biogas

Electrical parameters

Analog input 2-channel 4-20 mA input

One way for external sensors, which can be displayed on the screen

Another way for dynamic compensation of process conditions

Analog output 2 channels 4-20 mA output

The first output range is consistent with the calibration range of the instrument

The second path is 0-100% user configurable

Alarm Two way Single Magnetic Dual Transformation (SPCO) Relay Numerical Alarm (250 V

5 A max)

Data collection users can collect the output signals of the analyzer through digital communication

The machine can store the maximum and minimum of 40 alarm points and oxygen content readings

Value, these data are all accompanied by date and time explanations

Digital Communication Modbus RTU Based on RS485 Protocol

Power supply 24 V DC; Maximum 1.5 A

The cable entry analyzer is equipped with 3 x M20 cable entries. Cable gland, conduit

The entrance and plug are provided as attachments

operating conditions

Environmental temperature 5~+55 ° C

Environmental pressure 750mbar~1250mbar

mechanical parameters

Preheating time<25 minutes

Stability time 5 minutes

Size 234 x 234 x 172mm (w x d x h)

Weight 9.7kg

Gas path material 316&430F stainless steel, silicone glass, platinum 3M 2216 adhesive

O-ring material: silicone, fluororubber or synthetic rubber

Gas connections EX1, EX3&GP2: 1/8 "NPT (cathode)

GP1 and GP3: 1/4 "NPT (cathode)

Protection level IP66, NEMA 4X

Hazardous Area Certification

ATEX II 2GD Ex d IIB+H2 T6 Gb

Ex tb IIIC T85°C Db IP66

IECEx Ex d IIB+H2 T6 Gb

Ex tb IIIC T85°C Db IP66

CCSAus Class I, Division 1, Groups B,C,D, T6

INMETRO Ex d IIB+H2 T6 Gb

Ex tb IIIC T85°C Db IP66

TC TR Ex 1 Ex d IIB+H2 T6 Gb

Weekly sealing ring

Under ATEX and IECEx environments

The temperature range

硅: Ta = -40°C~ +55°C

Fluororubber: Ta=-15 ° C~+55 ° C

Synthetic rubber: Ta=-10 ° C~+55 ° C