High temperature 260 ℃ pipeline insertion type temperature and humidity sensor

NegotiableUpdate on 05/01
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

High temperature 260 ℃ pipeline insertion type temperature and humidity sensor

Product Details

0to260℃ high temperature typeWM-RH/T -JtypePipeline typeinsertentertypetemperature and humidity sensor

1overview

The control of humidity in industrial processes is becoming increasingly strict, and our company's resistive humidity sensor, with its * * temperature and humidity resistance, can-70—260A wide temperature range of ℃ and1%---98%RHWithin the humidity range, it ensures high precision, long-term stability, service life, and superior repeatability for temperature and humidity measurement in process environments, becoming the most cost-effective product in this field. The importance of temperature and humidity measurement in industrial production and experimental processes is highlighted in the areas of high-temperature drying and aging of products, low-temperature and high humidity experiments, and storage. Widely used for temperature and humidity measurement and control in industries such as dryers, kilns, meteorology, environmental protection, cement maintenance, electricity, desulfurizers, pharmaceuticals, textiles, chemicals, metallurgy, feed, food, batteries, low-temperature and high humidity storage, wood, and wine industry.

2、 Technical parameters

offerings Electricity:DC24V

measure Cheng: Humidity:0-100%RH

Temperature: 0-260 ℃ (optional range)

Working temperature:-10-50(transmitter)

Temperature accuracy: ± 0.5 ℃(-20-100℃)

Humidity accuracy: ±3%RH±2%RH

15%-95%RH,25℃)

Response time: ≤2s

Output signal; Current type:4-20mA(Two wire system)

4-20mA(Three wire system)

voltage source0—10V

05V

Load capacity: Current type ≤600Ω

Voltage type ≥2KΩ

outside Shell:ABSwhite100mm×68mm×50mm (transmitter)

IIIAppearance, wiring



remarks:In this dimension diagramLLength not less than110mm.(customized400mm,Equipped with stainless steel flange

Instructions for wiring terminals:Any incorrect wiring may cause irreversible damage to the transmitter



Circuit board in plastic shell

24V: Power supply+

GND; Power supply-The two-wire transmitter does not have this terminal,

Three wire current output,,This terminal is between the 24V terminal and the RH terminal

RT: Temperature output

RHHumidity output

In the legendABTo display instruments, actuators, or acquisition cards,ACorresponding temperature output of the circuit,BThe road corresponds to humidity output.

Two wire current output



Three wire current output:



Three-wireVoltage Output


4、 Installation steps:

1Probe usedM27*2The installation thread or outer diameter is68mmThe plastic or cast aluminum movable flange is fixed on the wall or pipeline (optional thread or flange).

2Connect the transmitter cable to the acquisition device.

Installation position

1The transmitter should be placed vertically as much as possible, and during installation, ensure that the sensor is on top of it

The font should be in the positive direction to ensure that the sensor is located below the transmitter;

2The installation height is the height of the human body sitting or the main environmental area that requires measurement

Installation precautions

1Avoid generating temperature in areas that are easy to transfer heat and directly affect the test area

Installation in poor areas may result in inaccurate temperature and humidity measurements.

2Install in a stable environment area. Avoid direct sunlight, stay away from windows, air conditioning, heating and other equipment, and avoid facing windows or doors directly.


Fiveuse

1Carefully check and ensure that the wiring is correct before connecting itDC24VWhen measuring the power supply with a multimeter, the corresponding current or voltage will be measured Value.

2If you want to disassemble the transmitter, you must first disconnect the power supply and then proceed with the disassembly.

3This transmitter is an indoor type, and water should be avoided from entering the inside of the transmitter to prevent damage.




6、 Precautions

1Please read this manual carefully before use to ensure correct wiring. Any incorrect wiring may cause irreversible damage to the transmitter.

2Avoid generating temperature in areas that are easy to transfer heat and directly affect the test area

Installation in poor areas may result in inaccurate temperature and humidity measurements.

3Prevent direct damage to the sensor from chemical reagents, oil, dust, etc. Do not use it for a long time in environments with condensation or extreme temperatures. Do not use cold Thermal shock.

7maintenance

Long term use of the transmitter may cause drift. To ensure measurement accuracy, it should be calibrated once a year.

8Transportation and storage

1Try to avoid vibration and handle the transmitter gently.

2Long term storage conditions:104020%RH60%RH.

9、 Open box inspection

1. After opening the packaging, check if the transmitter is intact.

2. One transmitter, one instruction manual, and one certificate of conformity.

10、 Fault analysis and troubleshooting

1. When simulating output, if the transmitter output is0, or if the output value is not within the range, please check if the wiring is correct or secure Solid.