High and low temperature alternating test chamberAlso known as a temperature chamber, temperature circulation chamberhigh-low temperature chamberHigh and low temperature aging chamber, programmable high and low temperature aging test chamber, high and low temperature cycling test chamber, etc. Used to simulate the temperature changes in the atmospheric environment. Inspect the performance indicators of components and materials of electronic and electrical products, automobiles, aerospace, marine weapons, medical equipment (drugs), universities, research institutions, and other related products under high and low temperature (alternating) cyclic changes.
High and low temperature alternating test chamberinBattery cell testingThe applications mainly involve the following aspects:
1. Temperature cycling test: In the high and low temperature aging test chamber, cycling tests under different temperature environments can simulate the high and low temperature environments of battery cells in actual use scenarios, in order to evaluate the performance of battery cells in different environments.
2. High temperature aging test: By conducting a high-temperature aging test on battery cells in a high-temperature test chamber for a period of time, the stability and safety of the cells under long-term high-temperature environments can be evaluated to predict their lifespan and performance degradation.
3. Low temperature discharge test: By conducting discharge tests on battery cells in a low-temperature test chamber, the capacity and performance changes of the cells in low-temperature environments can be evaluated.
4. Constant temperature test: For some battery cells that need to work under specific temperature conditions, such as power battery cells, constant temperature tests can be conducted in high and low temperature aging test chambers to evaluate their performance in specific temperature environments.
Through relevant tests, the performance and stability of battery cells can be comprehensively evaluated, providing reference for the design and production of battery cells.

Main Configuration
•Temperature range: A: -20 ° C to 150 ° C B: -40 ° C to 150 ° C C C: -60-150 ° C
• Humidity range: 20% relative humidity to 98% relative humidity
• Display instrument:7-inch TFT color LCD display(Can be connected)RMCS Control Software
• Operation mode: Fixed value mode, program mode (preset 100 sets, 100 steps, 999 cycles)
• Control mode:BTC Balanced Temperature Control Method+DCC(Intelligent cooling capacity control)+DEC (Smart Electrical Control) (Temperature Testing Equipment)
BTHC balanced temperature and humidity control mode+DCC(Intelligent cooling capacity control)+DEC (Smart Electrical Control) (Temperature and Humidity Testing Equipment)
•Curve recording function:Equipped with battery protectionRAM, Can save the set values, sampling values, and sampling time of the device; The maximum recording time is 350skyWhen the sampling period is 1/At the minute)
•Software usage environment:Compatible with upper computer operating softwareXP, Win7, Win8, Win10 operating systems (provided by the user)
•Communication function: RS-485interfaceMODBUS RTUCommunication protocol andEthernet interfaceTCP/IPChoose one of the two communication protocols; Support secondary development
Provide upper computer operation software,RS-485The interface is linked to a single device, and the Ethernet interface can enable remote communication and connection of multiple devices.
•working method: A/B: Mechanical single-stage compression refrigeration system C: Double stage cascade compressorCooling method
Observation method: Heated observation window with LED internal lighting
Temperature and humidity sensing method:temperaturePT100 armored thermocouple
Humidity:PT100 armored thermocouple
Dry wet bulb thermometer method (only works in tests with humidity control)
• Security protection:Fault alarm and cause, processing prompt function、A fail-safe mechanism、Upper and lower temperature protection function、Calendar timing function(Automatic start and automatic stop operation)、self-diagnostic function
• Verification configuration: Access hole with silicone plug (50mm, 80mm, 100mm left)
Data interface: Ethernet+software, USB data export, 0-40MA signal output
execution standard
GB/T5170.2-2008 Temperature testing equipment
GB/2423.1-2008 (IEC60068-2-1:2007) Low temperature test method AB.
GB/2423.2-2008 (IEC60068-2-2:2007) High temperature test method BA. GB/T5170.5-2008 Damp heat test equipment.
GJBl50. 3(MIL-STD-810D) High temperature test method.
GJBl50.4 (MIL-STD-810D) Low temperature test method.
GB2423.3-93 (IEC68-2-3) Test Ca:Constant humidity and heat test method.
GB2423.4-93 (IEC68-2-30) Test Db:Alternating damp heat test method.
Equipment security protection
refrigeration system: Compressor overpressure protection
Compressor motor overheatingprotect
Compressor motor overcurrentprotect
Humidification system: Humidifier tube dry burning protection
Abnormal water supply protection
Abnormal water shortage protection
Test box protection:Adjustable experimentspaceovertemperature protection
Extreme overheating of air conditioning channelprotect
Internal setting of ultra-high temperature protection in the controller
Fan motor overheatingprotect
Other protections: Main power phase sequence and phase loss protection
Whole machine leakage protection
Load short-circuit protection
Mine protection