Welcome Customer !

Membership

Help

Nanjing Hilbert Instrument Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

High and low temperature alternating test chamber

NegotiableUpdate on 12/28
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
High and low temperature alternating test chamber, 150L, model: GDW-150, manufacturer of high and low temperature testing machine - Nanjing Hilbert, suitable for reliability testing of industrial products at high and low temperatures. The components and materials of electronic and electrical products, automobiles and motorcycles, aerospace, marine weapons, universities, research institutions and other related products undergo high temperature and low temperature (alternating) cyclic changes
Product Details

High and low temperature alternating test chamberProduction testing method:

According to the experimental methods and industry standards, it can be divided into alternating test and constant temperature test. Both test methods are upgraded and expanded on the basis of high and low temperature test chambers. Alternating test chambers refer to instruments that can set the required high and low temperatures, as well as the required temperature time, in one go. The test chamber will follow the set program, which is to set a fixed temperature to make the test effect closer to natural climate, simulate more severe natural climate, and thus increase the reliability of the tested sample.

High and low temperature alternating test chamberThe selection of production environment and reliability testing equipment should follow the reproducibility of environmental conditions:

It is impossible to fully reproduce the environmental conditions that exist in nature in the laboratory. However, within a certain tolerance range, people can accurately and approximately simulate the external environmental conditions that engineering products undergo during use, storage, transportation, and other processes. This statement can be summarized in the language of engineering, which means that "the environmental conditions (including platform environment) created by the testing equipment around the tested product should meet the requirements of the environmental conditions and their tolerances specified in the product testing specifications. The temperature chamber used for military product testing should not only meet the high and low temperature test values and test times specified for different types of products in the national military standards GJB150.3-86 and GJB150.4-86, but also meet the requirements for temperature field uniformity and temperature control accuracy in the test specifications. Only in this way can the reproducibility of environmental conditions be ensured in environmental testing.

High and low temperature alternating test chamberMain technical parameters and performance of production:

Studio size (mm)

GDW-50 GDW-100 GDW-150 GDW-225
300×400×450 400×450×550 500×600×500 500×600×750

Large high and low temperature test chamber

GDW-408 GDW-800 GDW-010 GDW-015
Studio size (mm) 600×800×850 800×1000×1000 1000×1000×1000 1000×1500×1000
Operation mode Fixed value operation, program operation
technology
parameter
temperature upper limit

+100℃/+150℃/ 200℃

lower temperature limit

0℃、-10℃、-20℃、-30℃、-40℃、-50℃、-60℃、-70℃、-80℃

temperature fluctuation ±0.5℃
Temperature Uniformity ≤2.0℃
temperature deviation ±2℃
Temperature controller for high and low temperature test chamber Imported temperature controller
Temperature sensor for high and low temperature test chamber Pt100 platinum resistor
refrigeration
system
Refrigeration compressor method

Imported fully enclosed compressor unit

Single stage compression refrigeration, binary cascade compression refrigeration

cooling method air cooling
Security protection measures Protection against overheating, overload, overvoltage, leakage, short circuit, etc
material Box/insulation material Cold rolled steel plate with electrostatic spraying for the outer shell, SUS304 stainless steel plate/ultra-fine glass wool for the inner liner
Standard Configuration One test hole of size 50, two sets of test shelves, and lighting fixtures