-Working principle of Hash Amtax Inter II ammonia nitrogen online monitoring analyzer in the United States:
Indigo phenol blue method (compliant with DIN 38406 E5 standard). Under the action of a catalyst, NH4+reacts with hypochlorite ions and salicylate ions in an alkaline medium with a pH of 12.6 to form indophenol compounds, which appear green. Within the measurement range of the instrument, the degree of color change is directly proportional to the concentration of NH4+in the sample. Therefore, by measuring the degree of color change, the concentration of NH4+in the sample can be calculated.
-Application industry:
The Amtax inter2 ammonia nitrogen online monitor is used for online monitoring of ammonia nitrogen concentration in drinking water and surface water.
-American Hash Amtax Inter II Ammonia Nitrogen Online Monitoring Analyzer Instrument Features:
Measurement of NH4+ion concentration in water using a dual beam and dual filter photometer. By measuring with a reference beam, the ammonia nitrogen online monitor eliminates the interference of factors such as turbidity and power supply fluctuations in the sample on the measurement results.
The measured values can be displayed in graphical or numerical form.
The ammonia nitrogen online monitoring device has functions such as automatic calibration and automatic cleaning.
● Built in refrigerator to ensure the storage temperature of reagents
Use the FILTRAX sampling preprocessing system for sample preprocessing.
● Data storage function, graphic display function
| Technical specifications of Hash Amtax Inter II ammonia nitrogen online monitoring analyzer in the United States | |
|---|---|
| Measurement range: | 0.02~2.00mg/L NH4-N |
| 0.1 ~ 20.0mg/L NH4-N | |
| 1.0 ~ 80mg/L NH4-N | |
| Accuracy: | ± 4% ± 0.02mg/L NH4-N of the measured value |
| (0.02-2.0mg/L NH4-N) | |
| ± 2% ± 0.02mg/L NH4-N of the measured value | |
| (0.1-80mg/L NH4-N) | |
| Sample flow rate requirements: | Minimum of 0.1 l/h |
| Measurement cycle: | 5, 10 minutes optional |
| Instrument calibration: | Users can choose between manual calibration or automatic calibration according to their needs. Maintenance cycle: 6 months |
| Display: | LCD screen user maintenance: generally 1 hour per month |
| Reagent consumption: | Chemical reagent A/B: 4-8 weeks, depending on the measurement interval |
| Zero point and standard solution: 12 months | |
| Cleaning solution: | 6-12 months, depending on the cleaning cycle. Analog output: 2 channels of 0/4-20mA, maximum load of 500 ohm. Alarm output: 2 relays, 24V/1A |
| Digital output: | MODBUS or PROFIBUS service port: RS232 |
| Environmental temperature: | 5~40℃ |
| Working power supply: | 230VAC ± 10%, 50Hz |
| Power consumption: | Approximately 310VA (including refrigerator) |
| ordering guide | |
|---|---|
| AmtaxTM inter2 ammonia nitrogen online analyzer, standard configuration includes: host, installation accessories, reagents for at least one month of use, operation manual, etc. It is recommended to be used in conjunction with Filtrax sampling and filtration system | |
| LPV397.99.01000 | 0.1~20mg/L |
| LPV397.99.01010 | 0.1~20mg/L, Equipped with MODBUS communication |
| LPV397.99.01020 | 0.1~20mg/L, Equipped with PROFIBUS DP communication |
| LPV397.99.21000 | 0.02~ 2mg/L |
| LPV397.99.21010 | 0.02 ~2mg/L, Equipped with MODBUS communication |
| LPV397.99.21020 | 0.02 ~2mg/L, Equipped with PROFIBUS DP communication |
| Reagents and consumables | |
| Range of 0-200mg/L: | |
| LCW802 | Chemical reagents (A, B and their respective additives) |
| LCW804 | Zero point standard solution |
| LCW862 | Standard solution (0.5mg/L NH4-N) |
| LCW819 | Cleaning solution (2 × 250mg/L) |
| LZV281 | Wear prone spare parts with a range of 0-20.00mg/L: |
| LCW802 | Chemical reagents (A, B and their respective additives) |
| LCW804 | Zero point standard solution |
| LCW803 | Standard solution (5mg/L NH4-N) |
| LCW819 | Cleaning solution (2 × 250mL) |
| LZV281 | Wear prone spare parts |
| FILTRAX sampling and filtering system (required for monitoring surface water or raw water from water plants) | |
| 57389-01 | Filtrax with 2m unheated sample pressure transmission line |
| 57390-01 | Filtrax with 10m heat tracing sample pressure transmission line |
| 57391-01 | Filtrax with 20m heat tracing sample pressure transmission line |
| LXV294.52.04000 | Filtrax with 30m heat tracing sample pressure transmission line |
| FILTRAX installation accessories: | |
| LZY714.99.43050 | Stainless steel filter membrane container pool side installation components |
| LZX676.00 | Controller installation components |
| FILTRAX spare parts and consumables | |
| LZX 677 | Filter membrane (1 piece set) |
| LZX 018 | Wear prone spare parts for one year of use |
| LZX 667 | Filtrax annual maintenance component |