1、 HSX-150 Constant Temperature and Humidity IncubatorTechnical Specifications:
1. Model: HSX-150
2. Capacity: 150L
3. Temperature control range and accuracy: 10-50 ℃, ± 0.5 ℃
4. Humidity control range and accuracy: 40-95% RH, ± 5% RH
IIHsx- 150Constant temperature and humidity incubatorInstructions for Use:
1. The HSX-150 constant temperature and humidity incubator should be placed in a dry, clean, well ventilated studio away from heat sources and sunlight, and placed steadily to prevent vibration and noise.
Before use, all control switches on the panel are in a non working state.
3. Place experimental samples on the incubator rack, and maintain an appropriate distance between each experimental bottle (or vessel) to facilitate the circulation of cold (hot) air convection.
4. Connect the power supply and place the power switch in the "ON" position. The indicator light will turn on, and the LED digital display will show the actual temperature inside the box.
5. Then follow the instructions for using the intelligent temperature controller to set the desired temperature and humidity values. The temperature controller will automatically control the temperature and humidity inside the box to remain constant (when the temperature inside the box is higher than the set value, the incubator will automatically cool down, otherwise the incubator will automatically heat up; the humidifier will automatically humidify).
IIIMaintenance and Warranty:
1.HSX-150 Constant Temperature and HumidityThe incubator should be reliably grounded.
To ensure effective heat dissipation during condensation, the distance between the condenser and the wall should be greater than 100mm. There should be a 50mm gap on the side of the box, and at least 30mm space at the top of the box.
3.HSX-150 Constant Temperature and HumidityDuring transportation, maintenance, and upkeep of the incubator, violent collisions should be avoided, and the maximum inclination should be less than 45 degrees.
After unplugging the power socket, insert it again after 5 minutes. If inserted immediately, the compressor components are prone to damage.
5. During the refrigeration operation of this incubator, it is not advisable to make the difference between the temperature inside the incubator and the ambient temperature greater than 25 degrees.
6. It is strictly prohibited to put volatile chemical solvents, explosive gases and combustible gases in the incubator. Combustible sprays are not allowed near the incubator to avoid ignition by electric sparks.
To prevent pollution, try to avoid condensation of frost on the inner wall of the studio when using at low temperatures.
When the refrigeration system is working, frequent opening of the door should be avoided, which is beneficial for maintaining temperature stability and preventing dust and dirt from entering. After the refrigeration system stops working, clean the working chamber with a soft cloth.