The HPC40 series handheld pressure calibrator is a pressure/mA circuit 2-in-1 calibrator with full temperature compensation from -20 to 50 ℃,pass throughRTD and CvD can obtain true temperature based on "true Ohm" technology!
1.Multi functional: pressure calibrator, loop calibrator, digital thermometer, differential pressure gauge, barometer
2.built-in1-2 pressure modules, optional atmospheric reference module. Multiple ranges can be combined, and gauge pressure/absolute pressure/differential pressure/atmospheric pressure can be switched freely
3.You can choose from vacuum toThe pressure range of 15000 psi/1000 bar allows for the selection of either a single sensor or various dual sensor forms
4.By addingBARO option, which can switch between measuring pressure and absolute pressure on the interval
5.By adding an externalAPM CPF series pressure module, one device can read three pressures
6.Range temperature compensation:-20~50℃
7.Accurate to the point of reading0.035%
8.Measure and output external circuit power or internalMA signal of 24 VDC loop power supply, with slope and step functions,55mA current measurement
9.Damping and resolution adjustable, measurement display more stable
10.Percentage range conversion, real-time calculation and display of relative error
11.5 intelligent storage records, with one click access to commonly used settings
12.Multiple parameters are displayed simultaneously, making calibration information clear at a glance
13.One handed operation, simple menu, all metal casing, color display screen
14.Calibration data at different temperature points, including uncertainty assessment
15.Flexible power supply mode, can be connected to an external power adapter, power bank, orUSB direct power supply
16. USB interface, can be directly connected to a computer, works with calibration software, and can also be used as a USB flash drive
1. Pressure range:
Gauge pressure(HPC41 single range):
300KPa,1MPa,3MPa,10MPa,30MPa,70MPa,100MPa
30Psi,100Psi,300Psi,1000Psi,3000Psi,10000Psi,15000Psi
3Bar,10Bar,30Bar,100Bar,300Bar,700Bar,1000Bar
Gauge pressure(HPC42 Dual Range): Any two ranges with the same unit mentioned above
Absolute pressure: optionalThe BARO module can convert gauge pressure into the corresponding absolute pressure range
Vacuum:Vacuum can be measured with a gauge pressure range below 3Mpa/300Psi/30Bar
2. gauge pressure/Absolute pressure accuracy:
18~28℃:±0.035%Rdg(30%~100%range),±0.01%F.S(0~30%range)
-20~50℃:±0.05%Rdg(30%~100%range),±0.015%F.S(0~30%range)
3. Vacuum accuracy:± 0.05% F.S (F.S=corresponding pressure range)
4. RARO atmospheric pressure reference module, range: 700~1100mBarA (10.153~15.194PsiA), accuracy: ± 0.00725Psi/0.5mBar
5. Current measuring range/Accuracy: 0~55.000mA/± (0.015% Rdg+0.002mA)
6. current outputrange/Accuracy:0~25.000mA/±(0.015%Rdg+0.002mA)
7. Voltage measurement range/Accuracy: 0~30.000V/± (0.015% Rdg+2mV)
8. temperature measurementrange/Accuracy:0~400Ω/±(0.015%Rdg+0.02Ω), -200~800℃, Pt100, 2-3-4 wires (℃, ℉, K, Ω)
9. Temperature compensation:-20~50℃
10. Storage temperature:-40~75℃
11. Switch test: dry contact/Less than 1K Ω, display open/close/dead zone values
12. Circuit power supply:24VDC ± 10%, maximum load 25mA
13. Display:320×240Pixel color graphic display
14. Power supply:3xAA alkaline battery/NiMH rechargeable battery, USB power supply/charging, dedicated adapter
15. Electrical interface:APM pressure module LEMO interface, 4mm current and voltage socket, temperature probe socket
16. Process connection:CPF internal thread
17. Communication interface:mini-USB
18. Shell material: extruded aluminum
19. Weight:689can