1、 Overview
The microcomputer controlled low-temperature constant temperature bath adopts single-chip microprocessor control, self-tuning PID adjustment, 7-inch LCD touch screen display, imported platinum resistance (Pt100) temperature measurement, high temperature control accuracy and small fluctuation; Compressors using fully enclosed technology (imported compressor units are used for series below -40 ℃ or 30 liters) have high refrigeration efficiency and low noise. The instrument works stably and reliably, with convenient and safe operation. This product has the following advantages:
1. Adopting microcomputer temperature control PID regulation and self-tuning control technology, the temperature control accuracy is high;
2. There is a sensor open circuit display function;
3. Has liquid level alarm function;
4. The mixing pump and other components inside the platform are all made of 304 stainless steel material;
5. Fully enclosed brand compressor, fast cooling;
6. There are folding handles on both sides for easy movement;
7. Fast, stable, and reliable temperature rise and fall;
8. The set temperature and measured temperature are displayed separately in two digital windows;
9. The circulation pump can establish a second temperature field by circulating a constant temperature liquid externally.
HP-TS series low-temperature constant temperature bathWidely applicable in fields such as biotechnology, medicine, food, chemical engineering, metallurgy, chemical analysis, petroleum, etc., it provides users with a high-precision, controlled, and temperature uniform constant field source.
2、 Main technical indicators
(6-liter series)
Model Name |
HP-TS0506 |
HP-TS1006 |
HP-TS2006 |
HP-TS3006 |
HP-TS4006 |
HP-TS6506 |
HP-TS8006 |
|
Temperature range (℃) |
-5~100 |
-10~100 |
-20~100 |
-30~100 |
-40~100 |
-65~100 |
-80~100 |
|
Temperature fluctuation (℃) |
± 0.02 (25 ℃, medium is water or alcohol) |
± 0.05 (25 ℃, medium is water or alcohol) |
||||||
Display resolution (℃) |
0.01 |
|||||||
Inner container size (L) |
6 |
|||||||
Opening size x depth |
150×150×150(mm) |
|||||||
Pump circulation mode |
Internal and external circulation |
|||||||
Pump flow rate (L/min) |
10 |
|||||||
Control method |
Microcomputer temperature control, self-tuning PID regulation |
|||||||
temperature sensor |
Platinum resistance Pt100 |
|||||||
Cooling method |
compressor |
Secondary compressor |
||||||
Operating Voltage |
220VAC±10%,50Hz |
|||||||
Total power (kW) |
1.2 |
3 |
||||||
Dimensions (mm) |
340×340×670 |
510×510×1100 |
||||||
Remarks |
Rectangular stainless steel inner liner, stainless steel tabletop, displaying settings and measurements separately |
|||||||
3、 Environmental requirements
The constant temperature bath should be placed on a sturdy workbench;
No vibration, no harmful gases;
The room temperature should be relatively stable and not fluctuate rapidly, otherwise it will affect the temperature control accuracy; The room temperature should be around 25 ℃, especially for constant temperature baths at -40 ℃, -65 ℃, and -80 ℃. Excessive room temperature can affect compressor heat dissipation, reduce refrigeration efficiency, and easily cause malfunctions;
The constant temperature bath should be regularly maintained, especially if the cooling rate slows down after working for a period of time. The plastic mesh plate of the constant temperature bath should be removed to remove dust from the compressor and heat dissipation plate.
4、 Precautions for use
1. Add liquid medium into the tank before use, and the liquid level of the medium should not be lower than 10mm of the worktable. A too low liquid level will damage the heater when powered on.
2. The selection of liquid medium in the tank should comply with the following principles:
a. When the working temperature is between 5-85 ℃, water is generally used as the liquid medium;
b. When the working temperature is between 85-95 ℃, a 15% glycerol aqueous solution can be used as the liquid medium; Oil should be used when approaching or reaching 100 ℃;
c. When the working temperature is below 5 ℃, alcohol or freezing liquid is generally used as the liquid medium.
3. After the cooling switch is turned off, the cooling switch should be turned on again with an interval of 15-25 minutes (depending on the ambient temperature);
4. Power supply: 220V 50Hz, the power supply should be greater than the total power of the instrument, and the power supply must have a good "grounding" device;
5. The instrument should be placed in a dry and ventilated area, with no obstacles within 300mm around the instrument;
6. When the working temperature of the constant temperature bath is low, attention should be paid not to put hands into the bath to prevent freezing damage;
7. After use, place all switches in the off state and cut off the power supply;
8. The instrument should be regularly cleaned to maintain the cleanliness of the workbench and operating panel;
5、 Malfunctions and troubleshooting
Serial Number |
Fault |
Reason |
Exclude |
1 |
No display |
The power supply is not connected properly Fuse damaged |
Check the reason for not connecting the power supply and reconnect it Replace the fuse, and if it burns out again, contact the manufacturer |
2 |
Do not heat |
Set temperature lower than actual temperature Heater damaged Poor or damaged contact of temperature sensor |
Reset temperature Contact the manufacturer for repair reporting Check the sensor connection wire and connect it properly, or report for repair |
3 |
display
|
Poor or damaged contact of temperature sensor |
Check the sensor connection wire and connect it properly, or report for repair |
4 |
Slow cooling speed |
Poor heat dissipation effect |
Remove the panel, clean the heat dissipation plate and compressor body |
5 |
Unable to reach the specified temperature |
Alcohol or frozen liquid contains too much moisture |
Replace with new alcohol or coolant |
6 |
Not cooling |
The fuse of the refrigeration system is damaged The cooling fan is not running Temperature system malfunction, frequent heating The compressor is not working |
Replace the fuse, and if it burns out again, contact the manufacturer Attention: After compressing the mechanism, there should be a 15 minute interval before opening Contact the manufacturer for repair reporting |