Measurement principle of HMA-TNI water quality heavy metal total nickel online analyzer:
The instrument adopts colorimetric spectrophotometry for detection, using high-temperature digestion of water samples and adding digestion reagents to the water samples. The sample is pumped to the high-temperature digestion unit. The total nickel in the sample is dissolved into divalent nickel ions. The divalent nickel ion forms an orange brown colored complex with dimethylglyoxime in an alkaline solution in the presence of an oxidant (ammonium persulfate). Perform spectrophotometric measurement at a wavelength of 470nm, and analyze the concentration of the sample by comparing the difference between the initial color of the sample and the color after adding the color developer.
-Typical applications:
Online monitoring of total nickel in surface water, industrial wastewater (such as nickel containing electroplating wastewater), and sewage plant inlet and outlet
HMA-TNI Water Quality Heavy Metal Total Nickel Online Analyzer Instrument Features:
Low waste liquid. The amount of waste liquid is only 1.9L/month
Low reagent consumption. No need to replace reagents for one month
● Automatic range, automatic dilution. Simultaneously possessing low range and high range modes
● Automatic backwash sample filter, automatically cleaned before and after each test
● Automatic calibration
5.7 "color LCD touch screen
● Chinese menu
| Technical indicators of HMA-TNI water quality heavy metal total nickel online analyzer | |
|---|---|
| Measurement method: | colorimetry |
| Measurement range: | 0.05-3 (low), 0.4-10mg/L (high) |
| Detection limit: | 0.01mg/l |
| Accuracy: | ±3%F.S |
| Repeatability: | ±3%F.S |
| Zero drift: | ±5%F.S |
| Measurement time: | 50min |
| Dissolving time: | 120 ℃ high temperature digestion |
| Automatic calibration cycle: | Closed, 1 day, 2 days... 7 days optional |
| Maximum reagent dosage: | 5L/month |
| Sample flow rate: | 1~3L/min, Take 50mL each time for measurement |
| Sample temperature range: | 10-40℃ |
| Signal output: | 4-20mA/RS232/RS485 |
| Installation environment: | Indoor installation |
| Power supply: | AC220V±10%,50Hz |
| Dimensions (height x depth x width): | 900 × 255 × 600mm (analyzer body) |
| Weight: | 60kg (without reagents) |
| ordering guide | |
|---|---|
| HMA-TNI total nickel online analyzer, standard configuration includes: host, reagent bottle, installation accessories, operation manual, and detection report. | |
| HMA-TNI | Total Nickel Heavy Metal Online Analyzer |
| Accessories Information | |
|---|---|
| 7321530K | Pure water tank (including liquid level sensor) |
| 7321540K | Waste liquid tank (including liquid level sensor) |