HIC anti hydrogen induced cracking hydrogen sulfide corrosion test deviceBasic Parameters |
|
|
specification name |
H2S-2000
|
Reactor capacity |
2300 ml |
external dimensions |
Φ140*150 mm |
All steel fume hood (anti-corrosion) |
850*1200*2350 mm |
Gas cylinder cabinet(4ps) |
1200*400*1100 mm |
Filter sedimentation tank |
800*800*800 mm |
Spray tower (optional) |
Φ500*1000mm |
Gas emission outlet |
Φ40 mm |
performance | |
2.1.Test environment conditions |
The ambient temperature is+25℃, relative humidity ≤85% |
2.2.test method |
GB/T 8650-2015 Evaluation Method for Hydrogen Induced Cracking Resistance of Pipeline Steel and Pressure Vessel Steel GB/T20972The Oil and Gas Industry Materials used in environments containing hydrogen sulfide in oil and gas extraction |
2.3.Temperature range of the test chamber |
24℃±3℃ |
2.4.Standard quantitative pump flow range |
Nitrogen:0300Ml/min accuracy 0.1ml hydrogen sulfide:0-300ml/minprecision0.1ml |
2.5.Leakage concentration probe detection range |
0-10ppmThere are two standard configurations, one designed inside the gas cylinder cabinet and the other designed to be linked with the touch screen in the ventilation area. Once a leak is detected, the equipment will immediately stop working and trigger an alarm. The leakage alarm value0.1-1ppm adjustable) |
2.6.Concentration control method |
Accurate import of quality and traffic |
2.7.Gas filtration method |
Bubble neutralization filtration+Activated carbon adsorption treatment |
3. Structural Introduction | |
3.1.material |
Reaction kettle:Made of transparent acrylic, pressure resistant0.2MPa; Link pipeline:Teflon material; fume hood:Steel plate spraying (anti-corrosion coating) trap:glass; beaker:glass; Waste disposal device:Transparent acrylic board |
3.2.Location of thermometer |
located inNear the equipment, the temperature near the reaction kettle can be continuously recorded |
3.3.seal |
Fluororubber sealing strip sealing, screw connection; |
3.4.intake method |
Steel cylinder type |
3.5.Exhaust gas interface |
φ40mm |
3.6.Composition of exhaust gas |
Weak alkaline high humidity gas, harmless to the five toxins,conform toGB16297-1996Industrial gas emission standards |
3.7.Device Type |
The inside of the reaction vessel is under positive pressure, with a pressure<0.1MPa(Optional anti-corrosion pressure detection system, once overpressure occurs, the equipment will shut down and emit a buzzing sound). |
3.8.Leakage alarm device |
becauseThe gas used in the experiment is hydrogen sulfide, which is toxic, so this equipment is designed with two setsLeakage alarm deviceOne of themLocated inside the gas safety cabinet,Another design is inside the fume hoodOnce there is a leak, an automatic sound and light alarm will be triggered, and the indoor air filter will be activated. The fan inside the air filter will start to extract air from the inside of the steel cylinder cabinet for replacement, and filter it before discharge until the monitoring probe cannot detect any corrosive gas inside. The machine will shut down to ensure the safety of the user. |
3.9.Gas Cabinet |
4PSGas cylinder cabinet, can be placed4bottle8LThe gas cylinder |
3.10.waste discharge device |
A waste discharge device is designed at the end of the internal process, mainly for configuring alkaline solution. The gas is bubbled by the alkaline solution and then adsorbed by activated carbon before being discharged into the factory's waste gas pipeline (if there is no waste gas pipeline in the factory, an additional waste gas treatment tower can be installed, and after three-stage filtration, it can be discharged outdoors) |
3.11.intake method |
The air intake of this device adopts the method of mass flow introduction, as shown in the attached test process for details |
3.12.Control Panel |
Programmable corrosion gas specific controller, power switch, paperless recorder |
|
Corrosion gas controller
|
The temperature and humidity control instruments are imported7.0-inch color touch screen programmable controller, with a large screen, easy operation, and easy program editing,beltR232communication port; 1. Accuracy:0.1 ℃ (display range); 2. Resolution:±0.1℃; 3. Temperature sensor:PT100 platinum resistance thermometer; 4. Control mode: thermal balance temperature and humidity control mode; 5. Temperature and humidity control adoptsCoordination control of the P.I.D+S.S.R system on the same channel; 6. It has the function of automatic calculation, which can immediately correct the temperature and humidity change conditions, making temperature and humidity control more accurate and stable; 7. The controller operation interface is available in both Chinese and English, and the real-time operation curve can be displayed on the screen; 8. have10 sets of programs, 100 segments per set, with a capacity to cycle 999 steps per segment, and a maximum time setting of 99 hours and 59 minutes per segment; 9. After inputting the data and experimental conditions, the controller has a screen lock function to avoid manual touch and shutdown; 10. haveRS-232 or RS-485 communication interface, which can be programmed on a computer to monitor the testing process and perform functions such as automatic power on/off, curve printing, and data processing; 11. The controller has an automatic screen saver function, which better protects the LCD screen (prolonging its lifespan) during long-term operation; 12. Accurate and stable control, long-term operation without drift; 13. 1s ~999hThemTheSThe spray stop time can be set arbitrarily; 14. The program or fixed value test can be set arbitrarily, and can be based onRequest the program to conduct a cyclic test; 15. instrument displayexperimentTemperature, andthree-channelgas flow rate(Each channel contains nitrogen flow and hydrogen sulfide flow); 16. All functions can be set and operated on this screen; 17. In operation or setting, if an error occurs, a warning signal will be provided; 19. The main control components are all Schneider electrical components from France, ensuring the reliable performance of the equipment. |
3.19 Protection System |
1Gas leakage alarm handling; 2Hazardous gas treatment and protection system; 3Exhaust fan overcurrent protection; |
4. transportation |
The test box is of integral type and can be transported as a whole |
5. Terms of Use |
The user guarantees the following conditions |
environment/Venuecondition |
ltemperature5℃~30 ℃ relative humidity: ≤85% atmospheric pressure86kPa~106kP lThe ground is flat, well ventilated, and free of flammable, explosive, corrosive gases and dust lThere is no strong electromagnetic radiation source nearby lThere is a drainage floor drain near the equipment (distance from the refrigeration unit)Within 2 meters lGround bearing capacity of the venue: not less than200kg/m2 lLeave appropriate maintenance space around the equipment lThe airflow around the equipment is unobstructed, without high concentration dust, corrosive or flammable gases lSuggest reserving independent storage space |
Power supply conditions |
lVoltage:AC220V/50HZ lAllowable voltage fluctuation range:AC(1±10%) lAllowable fluctuation range of frequency:(1±1%)50Hz lThe grounding resistance of the protective grounding wire is less than4Ω; TN-S power supply or TT power supply lRequire users to configure corresponding capacity air or power switches for the equipment at the installation site, lAnd this switch must be independent and specifically designed for use with this device lTotal maximum power:1000W |
6. Precautions |
1. Due to the strong adsorption of hydrogen sulfide, it is necessary to wear a specialized gas mask and rubber gloves when operating this equipment; 2. Before entering the laboratory each time, it is necessary to observe whether the three color lights that come with the equipment are in green state 3. Nitrogen pressure holding test must be conducted during each test before conducting deoxygenation and hydrogen sulfide tests |
|
HIC anti hydrogen induced cracking hydrogen sulfide corrosion test deviceThe equipment mainly consists of the following parts: 1 Reaction kettleThree sets (standard configuration) 2 trafficThree sets of experimental equipment (standard configuration) 3 Leakage alarm deviceTwo sets (standard configuration) 4 One anti-corrosion all steel fume hood (standard configuration) 5 Two traps (standard configuration) 6 Two beakers (standard configuration) 7 Exhaust gas filtration device (standard configuration) 8 8L hydrogen sulfide gas bottle (standard configuration, equipped with dedicated pressure reducing valve) 9 8L nitrogen gas bottle (standard configuration, equipped with a dedicated pressure reducing valve) 10 Filter consumables (*500g two bottles, 2kg large particle activated carbon) 11 sprayFilter tower(optional) 12 Electronic scale(optional) 13 Gas(Optional, except for equipment configuration) 14 backupPressure reducing valve (optional) 15 Alarm (optional, in addition to standard) | |