-Measurement principle:
The NH6000sc analyzer uses gas sensing electrodes (GSE) to measure ammonium concentration. The ammonium ions in the sample are first converted into gaseous ammonia. Only ammonia gas passes through the gas permeable membrane of the electrode and is detected. This method supports a wide measurement range and is less susceptible to cross sensitivity than methods based on ion selective electrodes (ISE).
-Application industry:
surface water
drinking water
Industrial production process water
Wastewater treatment process
-Instrument features:
●Experience excellent efficiency and reliability: NH6000sc ammonia nitrogen analyzer provides precise measurement and predictive diagnosis, demonstrating stability and reliability
Simplify your online operation with the NH6000sc online ammonia nitrogen analyzer, which uses gas sensing electrode (GSE) technology. The environment controllable design prepares for simple outdoor installation at the measurement points you need. The Hach NH6000sc online gas sensitive electrode ammonia nitrogen analyzer, with twice yearly maintenance requirements and verified reliability, will provide you with reliable measurement results to improve process efficiency.
●Improve uptime and accuracy through reliable technology
Hach's NH6000sc ammonia nitrogen analyzer uses fast, accurate, and reliable GSE technology. NH6000sc provides reliable measurement with automatic calibration, verification, and cleaning functions. The consistency between online data and laboratory data has been improved by automatically sampling and validating the captured samples. The predictive diagnostic and supportive maintenance workflow helps you extend your normal operating time.
●Durable and lightweight integrated filtering options
The integrated FX610 filtration system and the more robust FX620 have a lightweight structure, improving air cleaning and flow detection functions. These features ensure reliable performance and reduce operational touchpoints. In addition, you can remotely monitor the operation of the analyzer to ensure stable and reliable measurement data.
●Maintenance every six months, saving time and effort
The NH6000sc ammonia nitrogen analyzer reduces the frequency of routine user maintenance to only twice a year. The instrument is designed with efficiency as its core and a streamlined structure, making reagent operation simple and convenient. In standard cleaning mode, the replacement time of reagents can be extended to once every six months.
| technical indicators | ||||
|---|---|---|---|---|
| Measurement Range 1 | Measurement range 2 | Measurement range 3 | Measurement range 4 | |
| scope | 0.02 - 5.0 mg/L NH4-N | 0.05 - 20 mg/L NH4-N | 1 - 100 mg/L NH4-N | 10 - 1000 mg/L NH4-N |
| Lower Limit of Detection (LOD) | 0.02 mg/L NH4-N | 0.05 mg/L NH4-N | 1 mg/L NH4-N | 10 mg/L NH4-N |
| precision | ≤ 1 mg/L: 3% measured value ± 0.02 mg/L >1 mg/L: 5% measured value ± 0.02 mg/L |
3% measurement value ± 0.05 mg/L | 3% measurement value ± 1.0 mg/L | 4.5% measurement value ± 10 mg/L |
| repetitiveness | 3% measurement value ± 0.02 mg/L | 2% measurement value ± 0.05 mg/L | 2% measurement value ± 1.0 mg/L | 2% measurement value ± 10 mg/L |
| response time | ≤ 0.2 mg/L: 3 measurements (minimum 15 minutes) >0.2 mg/L: 1 measurement (5 minutes) |
<5 minutes | <5 minutes | <5 minutes |
| measurement method | Gas sensing electrode (GSE) | |||
| measurement interval | 5, 10, 15, 20 or 30 minutes (user adjustable) | |||
| PH range | 5 - 9 pH | |||
| Permissible range of chloride ions | ≤ 1000 mg/L Cl - | |||
| sample quality | Ultrafiltration or similar (FX6x0<0.45 μ m) | |||
| Sample flow rate | 1.0-20.0 L/h to analyzer, filtered from external samples | |||
| Sample pressure | Maximum 2.5 bar (0.25 MPa) (36.3 psi) | |||
| Sample temperature | 4 - 40 ℃ (39 - 104 ℉ | |||
| Special Remarks | The liquid level in the pool must be lower than the bottom of the analyzer | |||
| Operating temperature range | -20 - 45 ℃ (-4 - 113 ℉ ); 95% relative humidity, no condensation | |||
| Storage conditions | -20 - 60 ℃ (-4 - 140 ℉ ); 95% relative humidity, no condensation | |||
| power supply | Powered by SC controller. When using SC4500, an additional power box is required. Analyzer and heated drain pipe (powered by analyzer): 115 VAC or 230 VAC |
|||
| cable length | Data and power cables: 2 meters (80 inches) from the edge of the casing | |||
| material | PUR 66 UV resistant | |||
| Shell grade | Grade: IP55, NEMA 3R UL50E Material: PUR 66 | |||
| Controller compatibility | Digital SC controller. The controller is sold separately. | |||
| size | 575 x 991 x 425 millimeters (22.63 x 39.01 x 16.73 inches) | |||
| weight | Approximately 45 kilograms (99.21 pounds), without reagents | |||