GPC gel purification can automatically complete sample injection, separation and purification, and target component collection. GPC Cleanup workstation performs data collection, storage, and management functions.
1Gpc gel purificationsystemconfiguration
P15High pressure infusion pump |
UV15Variable wavelength ultraviolet detector |
7725iManual injection valve |
Gel purification column chromatography (medium pressure glass column tube400*25mm, fillersBio-Beads S-X3) |
Chromatography workstation |
Automatic fraction collector |
Instrument spare parts package |
IIGpc gel purificationTechnical Specifications:
1) P15High pressure infusion pump
Flow range:0.001ml/min~10ml/min
Traffic accuracy:≤0.5%
Flow stability: RSD<0.1%
Pressure range: 0.00MPa~42MPa (420Kgf/cm)2)
Imported motor, imported driver, segmented into 20000 steps, no mechanical noise; Ensure accuracy.
Imported sealing system with strong high pressure resistance.
Imported high-efficiency pulsation damper with strong buffering capacity to minimize pulsation.
Display mode: LCD screen display.
Display mode: actual pressure, flow rate setting, pressure upper and lower limit setting.
Safety control: pressure upper and lower limit control, buzzer alarm, automatic shutdown.
2)UV15 variable wavelength ultraviolet detector
Optical system: Imported design diffraction grating monochromator, imported photoelectric receiving device.
Light source: Hamamatsu L6302 deuterium lamp from Japan
Wavelength range: 190-700nm
Wavelength accuracy: ± 1nm
Wavelength repeatability: ± 0.5nm
Noise: 2.5 × 10-5AU (p-p) (dynamic index, mobile phase methanol, 254nm, column attached)
Drift: 1 × 10-3AU/hr (dynamic index, mobile phase methanol, 254nm, column connection)
Attenuation range: 0.002-2AU/mV, 10 levels
Detection pool volume: 4.5Ml;
Small detection limit: 2 × 10-8G/ml (naphthalene/methanol solution)
LCD screen display, electrically controlled wavelength finding, electrically controlled zero adjustment, simple and easy to learn.
*The deuterium lamp is controlled by a starting voltage, which allows the deuterium lamp to have a lifespan of up to 2500 hours.
The design of the counter pressure device makes it difficult for bubbles to enter the circulation pool.
3)Workstation
It has batch processing function and can perform batch table editing, insertion, deletion, saving, pause and other functions. Data collection, data calibration, data processing and reporting, data output toWordorExcelFurther development in the middle
4)Automatic fraction collector
Display mode: LCD Chinese blue screen display
Collect test tubes: 100 tubes (each tube has a large capacity of 12ml (with a waste liquid tank)
Timing range: 1 second -999 hours, 59 minutes and 59 seconds, can be selected freely
Drop range: 1 drop -9999 drops, choose freely
Peak range (Type B): 1-20mv, choose freely
Programmable range: 1-10 segments can be selected freely
Movement mode: Hanging X-Y movement
Tube replacement control: 12V DC solenoid valve
Operation mode: Full Chinese menu style
Number of programs: 1-10 per mode
Power outage data storage: ten years
Test tube tray rack: square design with waste liquid tank, free combination, optional