Gongming Xenon Lamp Illumination Test Chamber Also known as anti-aging testing machine,It is a machine that simulates and reproduces the real environmental climate for industrial product testing. Currently, the two best-selling products produced by LiuqinClimate resistance test equipmentThere are xenon lamp aging test chamber and ultraviolet aging test chamber respectively,The experimental steps or functional conditions of xenon lamp aging test chamber and ultraviolet aging test chamber may be similar, but their names guarantee that they must be different. The biggest difference lies in their working principles! Xenon lamp aging test chamber simulates all spectra of sunlight, includingUVUltraviolet radiationIRInfrared and visible light, xenon lamp spectrum295nmto800nmThe range is basically consistent with the spectrum of sunlight. The UV aging test chamber cannot simulate full spectrum sunlight. Its working principle is that for durable materials exposed outdoors, the short wavelength of ultraviolet radiation300~400nmIt is the main cause of aging damage. In the short wave region of ultraviolet radiation, i.e. from365nmIn the low wavelength range of sunlight, the UV aging test chamber can simulate sunlight well, but it is powerless for longer wavelengths. That's why both areClimate resistant test chamber|The reason why the UV aging test chamber of the anti-aging testing machine is more economical than the xenon lamp aging test chamber.
Gongming Xenon Lamp Illumination Test Chamber |Aging resistance testing machine--Xenon lamp aging test chamber parameters:
1Power supply: three-phase five wire380V/50Hz
2Xenon lamp power includes:3KWThe4.5KWThe6KWWait, different sizes and irradiation intensities will be matched with lamps of different powers.
3Blackboard thermometer:100±3℃
4Temperature range: room temperature+10℃~80℃
5Temperature fluctuation: ±0.5℃
6Temperature uniformity:±2℃
7Humidity range:40~80%
8Humidity fluctuation:±3%
9Humidity uniformity:±3%
10Distance between sample holder and xenon lamp350~782mm(Adjustable), water-cooled is a rotating hanging sample holder.
11Rainfall time0~9999minute,Continuously adjustable
12Light cycle:0~99hour99Minutes, interval (off) adjustable lighting
13Studio size: available700×600×600; or800×800×800(Depth x Width x Height) More sizes can be customized.
Aging resistance testing machine--UV aging test chamber parameters:
1Radiation intensity: low: 0.35 W/m2 (UVA, UVB)tall: 1.4 W/m2 (UVA), 1.23 W/m2 (UVB)
2Radiation uniformity: covering which surface of the sample4%
3Blackboard temperature range: UV stage -BPT 35-80°C (95-176°F)Condensation stage -BPT 35-60°C (95-140°F)
4Sample capacity:48a+BPTin24A sample clip or placement1140×640×690(DWidth xWDeep xHOn the high sample compartment.
5Weight:150KGLeft or right, depending on the selection of accessories
6Power supply: Voltage220V±10%Single phase three wire system,50/60Hz, current8A
7Spray water pressure:25-40psi(1.7-2.7bar)Purity:<1 ppmDissolved solids, silicon:<0.1 ppmYes, conductivity:<5 us/ cmor200Thousand Euro or Reference Test Method
8Requirements for condensate water:Stress2-60psi(0.1-4.1bra)Purity: Deionized water is recommended, but not necessary.