DSC-500CGlass transition temperature tester
Differential Scanning Calorimetry (DSC), as a classic thermal analysis method for thermal effects under controllable program temperature, has been widely used in various fields such as research and development, process optimization, quality control, and failure analysis in materials and chemistry.
By using DSC method, we can study the phase transition of inorganic materials, the melting and crystallization process of polymer materials, the polymorphic phenomenon of drugs, and the solid/liquid phase ratio of foods such as oils and fats.
Glass transition temperature testerCompliant with standards:
ASTM E794-2001 Test Method for Melting Point Temperature and Crystallization Temperature for Thermal Analysis
ASTM E793-2001 Test Method for Determining the Heat of Melting and Crystallization Processes by Differential Scanning Calorimetry ASTM D3418-2003 Test Method for Determining the Transition Temperature of Polymers by Thermal Analysis
ASTM D3417-1999 Test Method for Determination of Melting Heat and Crystallization Heat of Polymers by Thermal Analysis
ASTM D3895-2003 Test Method for Oxidation Induction Time of Polyolefins by Differential Scanning Calorimetry ASTM E537-2002 Test Method for Thermal Stability of Chemicals by Differential Thermal Analysis ASTM E487-1999 Test Method for Constant Temperature Stability of Chemical Materials
JIS K7121 Determination Method for Plastic Transition Temperature
JIS K7122 Determination Method for Transition Heat of Plastics JIS K7123 Determination Method for Specific Heat of Plastics
GB/T13464-1992 Determination of Thermal Stability of Materials by Thermal Analysis DIN H7101-1989
ISO11357-3 Determination of Crystallization Temperature and Crystallization Enthalpy Change of PE by DSC
Main Features
l Easy to operate, no testing experience required, only a small amount of training needed
l The software is suitable for computer screens of various resolutions
l Dual temperature probes ensure high precision and repeatability
l The experimental process does not require personnel supervision
l Digital gas mass flowmeter automatically switches between two gas channelsbodytraffic
l Software can be upgraded online for free
l Seven inch large screen LCD display, with clear graphs and curves at a glance
l Built in high sensitivity differential thermal sensor imported from abroad, with higher accuracy, capable of measuring weak signals(DSC-500C)
Technical parameters
DSC range |
0~±500mW |
temperature range |
room temperature~500℃ |
|
heating rate |
0.1~50℃/min |
Temperature resolution |
0.01℃ |
|
temperature accuracy |
±0.1℃ |
Temperature repeatability |
±0.1℃ |
|
DSC accuracy |
±2% |
DSC resolution |
0.001mW |
|
DSC resolution |
0.001mW |
temperature control |
Full process automatic control |
|
Curve scanning |
Heating up scan |
Atmosphere control |
Gas mass flowmeter |
|
display mode |
24 bit color, 7-inch touch screen |
working power supply |
AC220V 50Hz/60Hz |
|
Size (w * d * h) |
46*35*27cm |
Net weight |
14.5KG |
|
Packaging size (w * d * h) |
58*45*40 |
Gross weight |
18KG |
|
|
|
Sensor DSC-500C |
high sensitivity |
|
|
Computer configuration requirements |
hardware configuration l Saiyang dual core l 2GB system memory l Resolution 1366 * 768 pixels or higher l Hard drive 500GB
|
software configuration l Win2000 or higher system l Adobe PDF Reader |
||
Application Examples
Measure physical and chemical changes related to heat, such as glass transition temperature and melting point、Melting temperature, crystallization and crystallization heat, heat of phase transition reaction, thermal stability of the product, solidification/crosslinking, oxidation induction period, reaction kinetics, specific heat, etc. Note: The thermal stability test during the oxidation induction period is applicable to the national standard GB/T 17 3 9 1-1 9 9 8.
Instrument Principle
Differential scanning calorimetry is a technique that measures the power difference between a substance and a reference substance as a function of temperature under programmed temperature control. The DSC-500 series differential scanning calorimeter mainly consists of a heating furnace, a main unit, a microvolt amplifier, an A/D converter, a data acquisition system, a gas flow control system, a computer, a printer, and other components, supplemented by the switching of two atmospheres. The measurement results are processed by a computer data processing system.
Software interface:

Glass transition temperature automatic calculation result interface
Enthalpy variable, epitaxial starting point, melting point calculation result interface
Enthalpy ratio calculation result interface
Calculation result interface for oxidation induction period
DSC-500 series accessories list
Serial Number |
Name |
quantity |
1 |
DSC-500 type heating furnace and main unit box |
1 |
2 |
Professional thermal analyzer software |
1 |
3 |
power cable |
1 |
4 |
signal cable |
1 |
5 |
Glass tube fuse (3A) |
4 |
6 |
tweezers |
1 |
7 |
Sample key |
1 |
8 |
Certificate of Conformity |
1 |
9 |
Ventilation connector |
4 |
10 |
Aluminum crucible (Φ 6.7 × 3mm) (Belongs to consumables, with discounted prices for future purchases)2.4 yuan per piece) |
400 |
11 |
Reference material( |
1g |
12 |
Standard samples (In, Sn, Zn) |
0.5g each |
13 |
DSC-500 Differential Scanning Calorimeter Manual |
1 |
14 |
Ventilation plastic pipe (blue) |
3 meters |
15 |
Ventilation plastic pipe (orange) |
3 meters |
16 |
softwareAdd MI dog |
1 |
Product images