Gas explosion simulation experimental device

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

This device belongs to the category of system development equipment. It mainly consists of six parts: the main body of the gas explosion combustion explosion container, the inlet and distribution system, the vacuum system, the ignition system, the circulation system, the safety device, and the container support. The equipment has reliable quality and high safety performance, and can intuitively display the process of gas explosion transmission in mine tunnels, as well as real-time display of changes in gas concentration in pipelines.

Product Details

1Gas explosion combustion explosion container body

Specification: Total length2.4Rice, combustion explosion cavity:≥φ130mmThe material is transparent quartz tube;

Maximum gas explosion limit pressure:≥1.0MPa

Sealing of combustion and explosion containers:50kPaUnder absolute pressure,5Minute pressure rise≤0.5kPa

Perform gas combustion explosion test after opening the pressure relief device at the end of the combustion explosion container.

2. Air distribution system: using partial pressure method for proportional air distribution

Pressure measurement range:0100 kPa(Absolute pressure), Resolution:0.1 kPaLCD display;

Response time:≤1s

Accuracy:±0.5%FS.

3. Vacuum pumping system

Rotary vane vacuum pump: vacuum degree can reach≥0.065MPa

Extraction rate: not less than2L/min.

4. Ignition system:

Ignition method: continuous pulse circuit discharge, discharge frequency100Hz

Ignition voltage:15 kV

火花能量:160J

Electrode spacing:18mm.

5. Circulating system: Adopting a closed oil-free circulation system, the gas and air in the combustion and explosion container are evenly mixed and reliably ignited;

A branch of the circulation system pipeline needs to be led out, with two tees reserved for two groups of students to learn how to operate and use the gas meter.

6. Safety device:

Adopting wireless remote control ignition, human-machine separation, remote control distance not less than15m

The safety interlock system cannot start the ignition system if the pressure relief device at the end of the combustion explosion container is not turned on.

7. Combustion and explosion container bracket

Use metal material brackets.