GDT-500 Ozone AnalyzerMeasure ozone concentration using the principle of ultraviolet absorption. Using the Beer Lambert law, measure the attenuation value of 254nm ultraviolet light passing through the sample using an ozone spectrophotometer. GDT-500 simultaneously measures sample gas and zero gas, performs real-time comparative calculations, and accurately measures ozone concentration.
Ø Built in low-noise and high-efficiency sampling pump;
Ø Built in gas flow and pressure monitoring functions, and supports status alarms;
Ø Built in zero gas source;
Ø The photometer using dual channel technology provides reliable measurement results;
Ø Data interface supportRS232/485,4~20mA- Oh.
| Main parameters | Parameter description |
| range | 0~100ppb,500ppb,1000ppb,20ppm,100ppm |
| Accuracy and precision | 1ppbor1%RD |
| traffic | 1 L/min |
| Zero point noise | 0.5 ppb (Avg 60s) |
| Low detection limit | 0.5 ppb |
| Zero drift(24h) | <1.0ppb |
| Range drift(30days) | <1%F.S |
| Precision | 1.0ppb |
| linear | 1%F.S |
| signal output | RS232/4~20mA |
| Power supply and power consumption | AC220V 30W |