It is a general term for devices that can generate a constant and known airflow or atmosphere of water vapor under certain conditions, and can be controlled within the range of relative humidity of 5-95% RH, while also achieving control of gas flow and temperature.
At the bottomEnvironmental temperature: 0~50 ℃
At the bottomRelative humidity: 10%~90% RH, non condensing
At the bottomInlet requirements: distilled water or deionized water (ultrapure water)
At the bottomAir intake requirements: Nitrogen or clean air, and non corrosive gas to 304 stainless steel
name |
Technical parameters |
Principle |
Dual temperature and dual pressure mixing method |
Humidity range(ppm) |
100~25000 |
Saturation pressure range(psig) |
0~150 |
Saturation temperature range(℃) |
0~70 |
Saturation heating rate |
2 Min/℃(Avg) |
Test pressure range(psig) |
0~50 |
Test temperature range(℃) |
0~100 |
Gas flow range(SLM) |
≤60 |
Gas flow accuracy |
≤±1.0% F.S |
Gas input pressure(psig) |
≤100 |
Gas type |
Zero gas or nitrogen gas |
System Display |
7”LCDLCD touch screen |
signal interface |
RS232/RS485/RJ45 |
Gas pipeline interface |
316L 1/4”Card sleeve |
Power supply |
220VAC@50Hz 500W |