- Phone
-
Address
No. 748, Group 7, Wudun Village, Lingang New Area, China (Shanghai) Pilot Free Trade Zone
Shanghai Juwei Instrument Equipment Co., Ltd
No. 748, Group 7, Wudun Village, Lingang New Area, China (Shanghai) Pilot Free Trade Zone
1:Name and Characteristics:
1.1.Product Name and Model:(JOOWAYx-ray)Energy dispersionXrayfluorescence spectrometer-------JW-EDX6000E
1.2.Manufacturer:Juwei Instrument Co., LtdLimited Company
fluorescence spectrometer
1.4. working conditions:
● Working temperature:15~30℃
● Relative humidity:40%~80%
Power supply: AC220V ± 5V
1.5.Product Features
1.5.1 EDX6000EIt can be specifically analyzed for the thickness of metal coatings.
1.5.2Breaking the traditional straight line design of instruments, adopting a streamlined integrated design, the instrument is simple and elegant.
1.5.3Adopting the latest technology from the United StatesSi-pinDetector, electrically cooled instead of liquid nitrogen cooled, with small size, accurate data analysis, and low maintenance cost.
1.5.4Adopting self-developed technologySESThe signal processing system effectively improves the sensitivity and accuracy of measurements.
1.5.5Seven types of optical path calibration and collimation systems, automatically switching according to different samples.
1.5.6Multiple anti radiation leakage design, radiation protection level belongs to one of the advanced products in the same category.
1.5.7The excellent integrated heat dissipation design greatly improves the overall heat dissipation performance and ensures the safe operation of core components.
1.5.8goodTemperature control technology for the movement ensuresXThe safe and reliable operation of the radiation source effectively extends its service life and reduces usage costs.
1.5.9Multiple instrument component protection system, and can be monitored throughout the entire process through software, making instrument operation more stable and safe.
1.5.10Specialized testing software, standard window design, user-friendly interface, and easy operation.
1.5.11adoptUSB2.0Interface, effectively ensuring accurate and high-speed transmission of data.
2.1.1. Si-pinElectric cooled semiconductor detector; resolution:149±5eV.
2.1.2.Amplification circuit module: for sample characteristicsXRadiation is used for detection, and the collected signals are further amplified.
2.2.1Filament current: maximum1mA.
2.2.2Belonging to semi consumable components,50Wair cooling.
2.3.1.Maximum voltage output:50kV.
2.3.2.minimum5kvControllable adjustment.
2.3.3.Comes with voltage overload protection.
2.4.1.Convert the collected analog signals into digital signals,And provide the processing results to the upper computer software.
2.4.2The maximum number of roads:4096.
2.4.3Includes signal enhancement processing.
2.5.1reduceXInterference during the transmission process of the ray path ensures accurate signal reception by the detector.
2.5.2Combine the collimator with the filter;
2.6.1Multiple options, with calibers of8、4、2、1The0.5mm.
2.7.1Free selection and switching of five types of filters.
2.8.1Up to dozens of collimators and filters can be freely combined.
3.1 Range of elemental analysis: from sulfur(S)To uranium(U).
3.2 Analysis detection limit can reach0.01Mm.
3.3 The analysis thickness is generally0.1Mmto30Mmbetween.
3.4 Repeatability can be achieved through multiple measurements0.1Mm(For less than)1MmThe outermost coating.
3.5 Long term job stability is0.1Mm(For less than)1MmThe outermost coating.
3.2.1Range of elemental analysis: from sulfur(S)To uranium(U)
3.2.2Can analyze simultaneously at once 24 An element
3.2.3Analysis detection limit can reach 100PPM
3.2.4The range of content analysis is generally 100ppm99.9%
3.2.5The software comes with multiple image processing and analysis calculation methods
3.2.6Repeatability can be achieved through multiple measurements 0.1%
3.3.1The range of element content analysis is2PPmto99.99%
3.3.2 Measurement time:60-200second
3.3.3 RoHSHazardous elements specified in the directive (limited)Cd/Pb/Cr/Hg/Br)Its detection limit is as high as2PPM
4.1 Sample cavity: Open large sample cavity.
4.2 Standard sample: Coating standard sample.
5.1 Free training for customer operators.
5.2 Installation, debugging, acceptance, training, and after-sales service are all provided free of charge at the user's site.
5.3Under normal use, if the malfunction is confirmed by our company to be caused by process or material defects and has not been disassembled or repaired, the instrument will be covered by a one-year warranty from the date of acceptance.
5.4 Lifetime product maintenance (customers must provide detailed and authentic valid purchase receipts, invoices, and warranty cards, etc.).
5.5 Free software upgrades are provided.
5.6 Provide effective technical services, and upon receiving user fault information,2Response within hours, if necessary,24Send someone to come and repair and troubleshoot within hours.
1.