Welcome Customer !

Membership

Help

Xu Changrui Taifeng Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Fixed bed reactor experimental apparatus

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The fixed bed reactor experimental device can be filled with 1-3ml of catalyst, the reaction heating furnace is tubular, the mass flow meter automatically controls and displays the gas flow rate, the liquid feeding is a micro level flow pump, the pressure is displayed digitally, and there is computer acquisition and control software that can collect system temperature, pressure, and liquid flow data. The system operation is stable, the data is reliable, and the repeatability is good.

Product Details

Fixed bed reactor experimental apparatus

Fixed bed reactor experimental apparatusDevice features

Reactor micro refillable1-3mlThe catalyst and reaction heating furnace are tubular, and the mass flow meter automatically controls and displays the gas flow rate. The liquid feeding is done by a micro level flow pump, with pressure digital display. There is computer acquisition and control software that can collect system temperature, pressure, and liquid flow data. The system operation is stable, the data is reliable, and the repeatability is good.

Fixed bed reactor experimental apparatustechnical indicators

1Operating pressure:0.2MPa.

2Usage temperature:600℃

3reactor DN10mm, Long550mmMaterial: Stainless Steel

4reactor 1Section heating heatpower1.5KW

5Temperature control instrument: digital instrument,0.25Grade accuracy,30Step program temperature control instrument:0.2Grade accuracy;

6Temperature and pressure display instruments: digital instruments,0.2Grade accuracy;

7pre-Heater : Diameter10mm, Long250mm

8pre-Hot furnace Heating power1KW

9gas-liquid separator : Diameter50mm, Long180mm

10dryer : Diameter25mm, Long400mm

11Gas flow rate:201000sccmQuality flow meter control and measurement

12Liquid flow rate:0.00110ml/minNon pulsating advection pump control.

13, SingleTCDDetector gas chromatograph and chromatographic workstation analysis system (with constant temperature valve box, automatic injection six way valve), gas chromatography online sampling and full analysis; Main technical specifications:

13.1Chromatographic column chamber

Column chamber size: width not less than260mmThe height is not less than280mmDepth not less than155mm

Temperature control range: under room temperature conditions Room temperature plus+3℃~400℃。

Temperature control accuracy: superior to±0.1℃。

Temperature gradient: room temperature plus+3℃~400℃, within the effective area of the column chamber, not exceeding1%.

The deviation between the set temperature and the indicated temperature shall not exceed1℃。

The deviation between the set temperature and the actual temperature shall not exceed2%.

Program heating order: greater than or equal to5Step by step.

Heating rate:0.140/min.

Set increment 0.1

13.2 Injection port vaporization chamber

Temperature control range: under room temperature conditions Room temperature plus+5℃~420℃。

Temperature control accuracy:±0.1℃ (room temperature)+15℃ to200Within ℃),200Above ℃ is ±0.2℃。

The deviation between the set temperature and the indicated temperature shall not exceed1℃。

Injection port program temperature rise: termination temperature400℃, heating rate: The heating rate is80/min,Set increment0.1;

Carrier gas pressure display: Carrier gas input pressure Electronic digital display; Pressure of carrier gas column head Electronic digital display

13.3 Thermal conductivity pool detector

Temperature control range: under room temperature conditions Room temperature plus+5℃~400℃。

Temperature control accuracy:±0.1℃ (room temperature)+15℃ to200Within ℃),200Above ℃ is ±0.2℃。

The deviation between the set temperature and the indicated temperature shall not exceed1℃。

Sensitivity: not less than4000mv·ml/mg(Benzene)

Noise: not greater than0.035mV

Drift: not greater than0.5mV/h

Automatic zero point correction, automatic bridge current addition, optional signal output channel.

13.4 External event control

The instrument can simultaneously set and control four external control events; Used for opening and closing ItoIVAn external event.

Multi stage control can be set, for example, when the external event is the injection switching valve, it can be used for various column switching operation control and valve injection operation. Valve start, reset, and multi-stage start, reset operations can be set.

13.5output signal

The instrument has a built-in chromatographic data sampling circuit, which can synchronously collect and output analog signals and digital data; Communication between chromatography and computer can be achieved, and the instrument's LCD screen can display spectra in real time .

14Computer control and data acquisition software control: can control the collection of system temperature, gas, liquid flow rate and other data, and collect pressure data.

External equipment:H2Steel cylinders and pressure reducers or hydrogen generators, N2Steel cylinders, pressure reducers, and computers.