- Phone
-
Address
6-2307 Tianshan Road, Wuxi City, Jiangsu Province
Wuxi Junzhan Instrument Co., Ltd
6-2307 Tianshan Road, Wuxi City, Jiangsu Province
Fischer XDLM237 X-ray Thickness Gauge Agent
FISCHERSCOPE X-RAY XDLM237 Core Positioning and Characteristics of Fischer X-ray Thickness Gauge
Fully automatic measurement: equipped with a programmable electric XY+Z-axis sample stage, supporting automatic scanning and multi-point measurement of batch parts.
Micro focusing high precision: The minimum spot diameter is 0.1 mm, designed specifically for small precision parts such as connectors and pins.
Top down measurement: suitable for curved surfaces and irregular parts, not limited by sample flatness.
Large sample compatibility: The body comes with a C-shaped groove, which can measure flat workpieces such as large-area PCBs.
Compliant with international standards: Measurements comply with DIN ISO 3497 and ASTM B 568.
Detailed technical parameters
1. Core system
X-ray source: Micro focused tungsten target tube, three levels of high voltage (10/30/50 kV)
Detector: Proportional Counter
Measurement elements: Aluminum (Al, 13) - Uranium (U, 92), can simultaneously analyze ≤ 24 elements
Measurement mode: single-layer/multi-layer coating thickness, alloy composition, hazardous substance (RoHS) screening
2. Optics and positioning
Collimator (4 types): 0.1 mm/0.3 mm/0.6 mm/0.5 × 0.15 mm slit
Filter (3 types): Nickel (Ni)/Aluminum (Al)/No filter
Positioning system: CCD color camera (40-160x zoom)+laser positioning+LED lighting
Distance compensation: DCM technology, automatically compensates for sample height differences
3. Sample stand (XDLM 237)
XY stroke: 255 × 235 mm
Z-axis travel: 140 mm
Moving speed: ≤ 80 mm/s
Repetition accuracy: XY ≤ 0.01 mm
Sample height: 140 mm
Load bearing capacity: standard 5 kg/20 kg (reduced accuracy)
4. Physical specifications
Size: 570 × 760 × 650 mm (W × D × H)
Weight: Approximately 120 kg
Power supply: 115/230 V, 50/60 Hz
Power consumption: 120 W