MF300Fm+is a new generation instrument developed by DIVERSE in the UK, which measures the ferrite content of austenitic stainless steel and duplex stainless steel welds based on the principle of magnetic field range. The sturdy ultra-thin probe benefits from a novel magnetic design that provides good coupling with the measured sample. This sensitivity result can well define the good response of samples with a full range of ferrite content. The new software supports calibration of the entire range of secondary ferrite samples according to NIST standards. A series of FN ferrite standard test blocks are provided, allowing users to calibrate and validate the main calibration. Store up to 1000 measurement data. You can choose to communicate with the computer through RS232 or USB, including downloading records or all or part of the database, as well as uploading record IDs. The record structure includes record numbers (0-999), measured ferrite numbers, and record IDs (4 characters, valid characters 0-9, A-Z, and spaces).
1、 Overview of the characteristics of MF300Fm+ferrite content measuring instrument:
1. Durable and portable
2. Standard or peak measurement of ferrite quantity
3. Estimation of Ferrite Percentage (Using FN to F% Curve)
4. Automatically reset to zero as needed
5. Probe selection: 10mm diameter probe or 4 × 6mm probe, each instrument can use up to 5 probes
6. Spring loaded sleeve (P46 probe)
7. Provide five conversion standards to check the accuracy of the instrument during operation
8. Can store 1000 sets of measurement data containing 4-character record IDs, imported into a computer using RS232/USB (optional)
9. Sampling at high temperatures (optional air-cooled probe -10mm diameter probe)
10. Simple measurement and small sampling area
11. Optional units (WRC-FN, 」)
12. Calibration standards can be traced back to the world's second standards
13. Provide transport boxes
2、 MF300Fm+Ferrite Content Measuring Instrument Probe:
1. Large area probe (P10, P10ac)
The large-area probe consists of a ring with a diameter of 10mm, which is set at the tip. For measurement purposes, this ring is in contact with a reasonable planar portion of the object being measured and can be read. The advantage of doing so is to average over a wide area, thereby eliminating the need for statistical analysis. Using an air flow adapter and other internal and calibration changes, the air-cooled probe version allows measurement of high-temperature samples. The air cooling is connected using standard hoses, passing through the sensor and magnetic group, and the exhaust vent is located on the probe side.
2. Small area probe (P10, P10ac)
The P46 small area probe (4 × 6mm) is still pen type, but the sensing area has been greatly reduced. It can be used to measure welds with a width less than 4mm. P46 has a spring-loaded sleeve to make the contact surface smaller. The advantage of the probe is its small sensing area, which is suitable for welding seams and pipelines. The end of the probe should be placed flat with the SUT.
The P46R probe has the same 4 × 6mm probe tip, but is configured to provide right angle sensing. Rectangular shapes facilitate operations in difficult to reach locations, such as inside pipelines.
3、 MF300Fm+Performance and Technical Parameters:
| unit | Quantity of ferriteFN
|
| Percentage of ferriteF%
|
| scope | 0-115FN |
| 0-80.5%(F%)
|
| resolution | 0.1FN |
| 0.1F% |
| Standard: Degree(10℃-30℃)
| |
| FN 0-10 | ±0.5 FN |
| FN 10-30 | ±5%reading
|
| | |
| FN 30-100 | ±10%reading
|
| Air cooling: degree(30℃-300℃)
| |
| FN 0-10 | ±1 FN |
| FN 10-30 | ±10%reading
|
| FN 30-100 | ±20%reading
|
| calibration | Factory calibration according toNISTstandard
|
| The user calibrates according to the provided standard test block |
| Provide more5Calibration (for)5Type probe)
|
| Reset to zero | On demand, automatic |
| pattern | Equilibrium or peak value |
| storage | 1000A belt4Characters (including0-9,A-ZRecord with spacesIDthe data
|
| Computer download and upload | Optional connection methods:RS232andUSB
|
| Function includes recording or partial/Download and record the entire databaseIDupload
|
| Probe P46 | Pen style, diameter15mm, Long130mm
|
| Sensing area4mm × 6mm(24mm2)
|
| Spring loaded sleeve |
| Probe P46R | Micro rectangle20×30×75mm
|
| Probe body90° Induction
|
| Probe P10 | Pen style, diameter15mm, Long120mm
|
| Sensing area diameter10mm(About)80mm2)
|
| Probe P10AC-Air cooling
| Pen style, andP10Same probe
|
| Adjust the air pipeline supply to reach1bar
|
| Optional oil filtration supply |
| air temperature:ambient temperature(10℃- 30℃)
|
| Operating temperature | standard0℃- 40℃(P46, P46RandP10)
|
| Air cooling0℃- 300℃(P10AC)
|
| humidity | frost-free |
| weight | 1.25kg |
| Transport box size | 210×100×45mm |
4、 MF300Fm+Ferrite Content Measuring Instrument Configuration:
1. Diverse MF300Fm+Ferrite Measuring Instrument Host
2. Ferrite probe/(choose according to requirements) Among them, ferrite analyzer probe (interchangeable, up to 5 optional):
(1) 4 × 6mm probe MF300Fm+P46
(2) 4 × 6mm right angle probe MF300Fm+P46R
(3) 10mm probe MF300Fm+P10
(4) 10mm probe with air cooling MF300Fm+P10ac
3. 5 standard test blocks
4. Transport box
5. Optional selection
(1) Serial/USB cable
(2) MF300+software services.