Explosion proof digital dust analyzer visual particle monitoring trend instrumentLD-8FThe explosion-proof digital dust detector is obtainedIIClass andIIIA fast dust analyzer with explosion-proof certification, which adopts the principle of light scattering and is equipped with a calibrated optical stability self calibration device. It can eliminate the influence of temperature and humidity drift and achieve rapid determination of the concentration of particles of different sizes in the air.Explosion proof digital dust analyzer visual particle monitoring trend instrumentSuitable for occupational health law enforcement personnel to detect dust concentration in the air of workplaces, and can also be used for non coal mines, tunnels, occupational health supervision, petrochemical industry, grain processing, transportation and storageConduct explosion-proof dust concentration testing at the production site of the enterprise.
JJG846-2015Measurement and Verification Regulations for Dust Concentration Measuring Instruments;
HJ653-2021《Environmental air particulate matter(PM10andPM2.5)Technical requirements and testing methods for continuous automatic monitoring system》;
GBZ2-2002Occupational Exposure Limits for Hazardous Factors in the Workplace;
GB3836.1-2021Explosive Environment1part:General Requirements for Equipment;
GB3836.4-2021Explosive Environment4part:By intrinsic safety type“i”Protected equipment ";
MT/T163-2019General Technical Conditions for Direct Reading Dust Concentration Measuring Instruments;
GB/T4208-2017The Protection Level of the Shell(IPcode)》;
GB/T18204.2-2014Methods for Health Inspection in Public Places2Part: Chemical Pollutants ";
GB/T18204.5-2013Methods for Health Inspection in Public Places5Part: Central Air Conditioning and Ventilation System
1.Detection principle: light scattering method, pump suction sampling method.
2.Measurement concentration range:(0.001~500)mg/m³ Detection sensitivity:0.001 mg/m³
3.Measurement time:The standard is1minute,1secondto9999secondCan be set arbitrarily
4.measurement accuracy:±10%;
5.Repetitive error:≤2%
6.Storage and data interface:Automatically store data while measuring,storage15000Group data,USBandRS232Interface
7.Operating temperature:-10~40 ℃
8.explosion-prooftype:Intrinsic safety explosion-proof, explosion-proof mark,Ex ib IIC T4 Gb; Ex ib IIIC T200130°C Db
9.Protection level:IP66
10.Working hours: Continuous work10hours or more
11.Overall dimensions:160mm×130mm×90mm
12.Weight:≤1.1kg